Operation
52
Imbalance display
If rotor imbalance is detected, shortly after the rotor
reaches 300 rpm, the message "bAL
" will appear in the
speed display.
The run is terminated, and
you may restart the centrifuge
after correcting the imbalance
(check loading).
Changing the settings during the run
You can change all settings during a run. By pressing
once any one of the "set" keys in the control panel you
can switch from the actual to the set point mode.
The setting to be adjusted flashes and can then be
altered. Once the data input is finished and the display
has changed to the actual value display mode, the new
settings become operative.
Stopping the centrifuge
Stopping with preset run time
Normally the run time has been selected, and all you
have to do is wait until the centrifuge terminates the run
automatically at the end of the set time.
As soon as the speed reaches zero, the display reads
"End". You can now open the centrifuge by pressing
the "open lid" key
and remove your samples.
If the lid has not been fully lifted out of the lid lock, the
message "Lift lid" appears (you must manually lift the
lid).
You can manually stop the centrifuge at any time by
pressing the "Stop“ key
.
At this point the remaining run time is displayed.
Stopping with continuous operation
If you have chosen continuous operation, you must
stop the centrifuge manually. Press the "Stop" key
in the control panel. The centrifuge starts deceleration
with the preset deceleration profile.
The display reads "End", and you can open the lid by
pressing the "Open lid" key
and remove your sam-
ples.
Содержание Sorvall Legend RT
Страница 1: ...SORVALL Legend T RT Instruction Manual QUIKset...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the 12 Notes...
Страница 44: ...Rotors and accessories 40 Notes...
Страница 60: ...Operation 56 Notes...
Страница 66: ...Maintenance and care 62 Notes...
Страница 82: ...78 Notes...
Страница 88: ...84...
Страница 100: ...96...