Troubleshooting
70
Error
Symptom
Possible causes and corrective measures
E-21
Instrument does not start or
brakes to standstill.
No rotor present or rotor identification impossible
A) Check whether a certified rotor is inserted.
B) Following a brief power failure, the rotor could not be
identified. Switch the instrument off and on again using
the main switch.
E-22
|
E-23
Rotor stops without brake
to standstill.
Instrument cannot be
operated.
Error in speed measurement
Switch the instrument off and on again. If the error persists,
contact KENDRO service
E-24
|
E-27
Instrument cannot be
operated.
Wrong status information from lid lock.
Switch the instrument off and on again. If the error persists,
contact KENDRO service.
E-28
Instrument decelerates
after a short starting-up.
Motor runs in wrong rotational direction.
Switch the instrument off and on again. If the error persists,
contact KENDRO service.