Troubleshooting
72
Error
Symptom
Possible causes and corrective measures
E-33
Rotor stops without brake
to standstill or does not
start.
Overpressure in the refrigeration system
1.
Turn instrument off and pull main plug.
1. Check and clean ventilation slots if necessary.
2. After about 60 min you can restart the instrument.
Observe the max. permissible environmental temperature!
If the error persists, contact KENDRO service.
E-34
Rotor stops without brake
to standstill or does not
start.
Overvoltage in the intermediate circuit.
Switch the instrument off and on again.
If the error persists, contact KENDRO service.
E-35
Rotor stops without brake
to standstill or does not
start.
Overcurrent in the intermediate circuit.
Switch the instrument off and on again. If the error persists,
contact KENDRO service.
E-36
|
E-38
Rotor stops without brake
to standstill.
Instrument cannot be
operated.
Overcurrent or error in current measurement.
Switch the instrument off and on again.
If the error persists, contact KENDRO service.
E-39
Rotor stops without brake
to standstill.
Instrument cannot be
operated.
Speed control measurement exceeds permissible rotor
speed.
Switch the instrument off and on again.
If the error persists, contact KENDRO service.
Содержание Sorvall Legend RT
Страница 1: ...SORVALL Legend T RT Instruction Manual QUIKset...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the 12 Notes...
Страница 44: ...Rotors and accessories 40 Notes...
Страница 60: ...Operation 56 Notes...
Страница 66: ...Maintenance and care 62 Notes...
Страница 82: ...78 Notes...
Страница 88: ...84...
Страница 100: ...96...