Removing the rotor
1. Open the centrifuge lid.
2. Remove the rotor cover (on applicable rotors).
3. Unscrew the clamping sleeve counterclockwise
using the socket wrench supplied with the instru-
ment until no resistance exists.
4. Grab the rotor with both hands and pull it perpen-
dicularly off the drive shaft carefully. Make sure not
to tilt it.
Grab rotor with both hands and pull
upwards perpendicularly .
When using an aerosol-tight bio-containment cover,
you may remove the respective rotor from the drive
shaft without opening the cover! You may then open
the rotor e.g. in a safety work bench and decontami-
nate it.
Audible alarm
Accompanying all error messages, a warning signal is
given out which only is silenced upon pressing any key.
You have the option of signaling the end of a run
acoustically. In order to activate respectively deactivate
this option please press the speed/rcf switch key . As
long as you keep this key pressed, the selection mode
is active.
The speed panel shows "beep“ and the time panel
shows "on“ or "off“.
By actuating the upward-key
in the time panel the
signal function can be switched on or off
.
When the message "rotor“ flashes, pressing
the start key once is sufficient to turn off the
warning signal and to accelerate the rotor to
the maximum speed displayed by the in-
strument.
Содержание Sorvall Legend RT
Страница 1: ...SORVALL Legend T RT Instruction Manual QUIKset...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the 12 Notes...
Страница 44: ...Rotors and accessories 40 Notes...
Страница 60: ...Operation 56 Notes...
Страница 66: ...Maintenance and care 62 Notes...
Страница 82: ...78 Notes...
Страница 88: ...84...
Страница 100: ...96...