Rotors and accessories
36
Aerosol-tight operation
Aerosol-tight rotors and tubes are only
to be opened in an approved safety
work bench when centrifuging danger-
ous samples!
Paying attention to the maximum per-
missible filling quantities is necessary!
Correct operation when filling the sample tubes and
closing the rotor cover are prerequisites for aerosol bio-
containment.
Before use, the seals in the rotors and
rotor covers, as well as the aerosol-
tight caps, have to be checked for
abrasion or damage and slightly
greased.
Replace damaged O-rings and seals!
Use the special lubricant
7600 3500 only to grease the seals !
Spare parts are delivered with the rotor or may be or-
dered separately.
Replace damaged or clouded caps
and lids of rotors and tubes
immediately.
The tubes are only to be filled such that the sample
does not reach the rim of the tube during centrifuging.