Troubleshooting
68
Error
Symptom
Possible causes and corrective measures
E-01
|
E-13
Rotor stops without
deceleration to standstill.
Instrument cannot be
operated.
Internal program error.
Switch the instrument off and on again. If the error persists,
contact KENDRO service.
E-14
Rotor stops with
deceleration to standstill.
Instrument cannot be
operated.
Overtemperature in the centrifuge tank.
Switch the instrument off and on again. If the error persists,
contact KENDRO service.
E-15
|
E-16
Rotor stops with
deceleration to standstill.
Instrument cannot be
operated.
Temperature measurement error.
Switch the instrument off and on again. If the problem
persists, contact KENDRO service.
E-17 Rotor
stops
with
deceleration to standstill.
Instrument cannot be
operated.
Max speed for the rotor identification exceeded.
Switch the instrument off and on again. If the problem
persists, contact KENDRO service.
E-18
Instrument stops with
deceleration to standstill
after short starting-up.
Bucket code for this rotor is not defined.
Check, whether the bucket is permitted for the inserted rotor
(rotor program, page 15).