Manual Operation
45
In ON-OFF operation with a fast rising and slowly falling flow a systematic positive error is to be expected. If the
rising is slow and the falling fast, the error will be negative.
WARNING!
If step response is used (e.g. for good reaction to fast changing flow), checking the accuracy for the
given application is strongly recommended!