Keito K7
Technical and maintenance
Manual
INT.022EN.02
Revision “I”
24/09/2010
Page 5 of 43
0 - System Adjustments. Group 0 of options.
0.0 - Memory reset.
This option resets the memory with the default values set at the factory:
- The default ticket header and footers are set.
- The default original inner height of the unit is set.
- The default original blood pressure adjustment values are set.
The other adjustments and parameters will not be modified. Steps to follow:
Key
Function to execute or select
Terminal Display
'Z' or 'ESC' Goes to option 0 of the main menu.
MAIN MENU:
0-SYSTEM ADJUSTMENTS
SPACE=NEXT
ENTER=SELECT
0
Select the system adjustments
SUB-MENU SYSTEM:
0.0-MEMORY RESET
SPACE=NEXT
ENTER=SELECT
0
Select the memory reset
ENTER=MEMORY RESET
OTHER KEY=CANCEL
ENTER
Starts the memory reset.
RESETTING THE
MEMORY
Once the reset is finished, the program goes back to option 0 of the
system adjustments sub-menu.
SUB-MENU SYSTEM:
0.0-MEMORY RESET
SPACE=NEXT
ENTER=SELECT
The memory reset has to be made in the following cases:
·
When you replace a motherboard.
·
Each time that you load the software in the motherboard.
0.1 - System information.
Option 1 of the system sub-menu, prints a ticket with the general status of the unit. Steps to follow:
Key
Function to execute or select
Terminal Display
'Z' or 'ESC' Goes to option 0 of the main menu.
MAIN MENU:
0-SYSTEM ADJUSTMENTS
SPACE=NEXT
ENTER=SELECT
0
Select the system adjustments
SUB-MENU SYSTEM:
0.0-MEMORY RESET
SPACE=NEXT
ENTER=SELECT
1
Select the printing of the system report. The printer starts the
printing of the ticket.
PRINTING THE
REPORT
Once the report is printed, the program goes back to option 1 of the
system adjustments sub-menu.
SUB-MENU SYSTEM:
0.1-SYSTEM REPORT
SPACE=NEXT
ENTER=SELECT