background image

 

 

Keito K7 

Technical and maintenance 

Manual 

 

INT.022EN.02 

Revision “I” 

24/09/2010 

Page 42 of 43 

 

Parameter #4 

The unit will send an SMS the day of the month indicated in this parameter 

Value  Description 

Will never send an SMS for this reason 

Sends an SMS with all data of status and collection and clears the collection counters the day 1 of the month 

Sends an SMS with all data of status and collection and clears the collection counters the day 2 of the month 

etc. 

 

 

Parameter #5 

The unit will send an SMS when the collection reaches the amount (in your local currency) indicated in this parameter 

Value  Description 

Will never send an SMS for this reason 

200 

Sends an SMS with all data of status and collection and clears the collection counters when collection is 200 ?? 
(your currency) 

etc. 

 

 

Parameter #6 

The unit will send an SMS when the number of issued tickets is equal or higher than the indicated in this parameter 

Value  Description 

Will never send an SMS for this reason 

200 

Sends an SMS with all data of status and collection and clears the collection counters when 200 tickets have been 
issued 

etc. 

 

 

Parameter #7 

Indicates the maximum time in tenths of second that the unit will wait to receive the confirmation that an SMS has been 
succesfully sent. The default value of this parameter is 1000 (10 seconds) 

Value  Description 

800 

The unit waits 8 seconds to receive the confirmation that the SMS was sent succesfully. 

 

 
 
 

Содержание K7

Страница 1: ...Keito K7 Technical and maintenance Manual INT 022EN 02 Revision I 24 09 2010 Page 1 of 43 MULTIFUNCTION K7 Technical Manual...

Страница 2: ...going to the computer 10 Switch the K7 on It is ready for use B Operating in the Advanced Programming Mode B 1 Description of the programming terminal The terminal has got an LCD screen with 4 lines o...

Страница 3: ...in menu if you press the SPACE key the options of the current sub menu will appear If when in a sub menu you press the ENTER key the program will execute the selected function If when in a sub menu yo...

Страница 4: ...Blood pressure measurement 1 Pressure calibration 2 Pressure continuous measurement 3 Pressure adjust 3 Escapement test 0 Impedance Test 1 Body fat calibration 4 Body fat and heart rate adjust 2 Heart...

Страница 5: ...tarts the memory reset RESETTING THE MEMORY Once the reset is finished the program goes back to option 0 of the system adjustments sub menu SUB MENU SYSTEM 0 0 MEMORY RESET SPACE NEXT ENTER SELECT The...

Страница 6: ...n if they don t belong to the pharmacy Prints the uses with smart card in the collection report Flag n 4 0 Format of the date DD MM YY Day Month Year 1 Format of the date MM DD YY Month Day Year 2 For...

Страница 7: ...x confirmation on the body fat measurement Prints the graphic of the heart rate analysis horizontally 4 Always measures the body fat 5 Body fat only in full operation The user must confirm the sex on...

Страница 8: ...sks once more Insert another coin 6 Once the minimum amount is reached the unit does not ask to insert more coins 7 Even if the minimum amount for use has been reached the unit asks once more Insert a...

Страница 9: ...cle disabled 6 The measure of the body fat is cancelled if the user steps off the platform without indicating the sex 7 Waits for the user to indicate the sex after the fat measure does not abort the...

Страница 10: ...flags The display show the uccrent value of the first flag INDICATE THE VALUE OF FLAG 1 3 AND PRESS ENTER ENTER Press ENTER until you get the flag number that you wish to modify in this example we wis...

Страница 11: ...ET SPACE NEXT ENTER SELECT 3 Select the programming of the price The display shows the price for the full operation currently programmed This price is correct in our example so you press ENTER to go t...

Страница 12: ...ight B press 5 Value in uros 9 0 Contabilized uros 9 4 Cummulated Weight and height 15 Weight height B press 10 Value in uros 16 0 Contabilized uros 16 4 Example of a collection ticket of a unit progr...

Страница 13: ...eport and the program goes back to the system adjustments sub menu SUB MENU SYSTEM 0 5 MAINTENANCE SPACE NEXT ENTER SELECT 0 6 Coin selector test This option allows the correct detection and operation...

Страница 14: ...Clear the counters The collection report has got two counters one is set to zero every time that a report is printed but the other one the cumulated keeps the total number of coins The only way to set...

Страница 15: ...Z or ESC Goes to option 0 of the main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENTER SELECT 1 Select the height system adjustments SUB MENU HEIGHT 1 0 RESET SPACE NEXT ENTER SELECT 0 Option 0 r...

Страница 16: ...mory when making a memory reset This adjustment will only be necessary if for any reason the inner height of the unit needs to be changed During the autotest process the unit makes an autozero of the...

Страница 17: ...the frequency is modified Let us suppose that we want to set the frequency to 52 29 kHz Proceed as follows Key Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main menu...

Страница 18: ...SC Goes to option 0 of the main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENTER SELECT 2 Select the weight system adjustments SUB MENU WEIGHT 2 0 RESET SPACE NEXT ENTER SELECT 0 Option 0 resets a...

Страница 19: ...tc since they can cause wrong measurements to the system Proceed as follows Key Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main menu Make sure that there isn t any...

Страница 20: ...steps The unit measures the value of the atmospheric pressure In this moment you must not apply any pressure to the sensor and the valve of the rubber bulb must be open A pressure of 200 mmHg is appl...

Страница 21: ...ometer and let the cuf deflate in full SUB MENU PRESSURE 3 1 CALIBRATION SPACE NEXT ENTER SELECT 3 2 Verify Pressure This option makes different continuous pressure measurements and shows them in the...

Страница 22: ...the decay rate 3 When the pressure is 50mmHg a ticket is printed with the elapsed time since the pressure was 250mmHg until reaching 50mmHg and with the maximum and minimum decay rate registered 4 Dur...

Страница 23: ...he main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENTER SELECT 4 Select the body fat adjustments SUB MENU FAT PULSE 4 0 IMPEDANCE TEST SPACE NEXT ENTER SELECT 0 This option measures the impedance...

Страница 24: ...mation In this case the electrodes record the electrical activity of the heart and from these signals the heart rate is measured At the end of the test the unit prints a ticket with the value of the p...

Страница 25: ...Keito K7 Technical and maintenance Manual INT 022EN 02 Revision I 24 09 2010 Page 25 of 43 Heart rate analysis Heart rate 86 1 P mn maximum Rate 90 P mn minimum Rate 83 P mn...

Страница 26: ...tial autotest and detects any malfunction Proceed as follows Key Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENT...

Страница 27: ...ADJUSTMENTS SPACE NEXT ENTER SELECT 5 Select the printer adjustments SUB MENU PRINTER 5 0 RESET SPACE NEXT ENTER SELECT 1 Select the paper feed Make sure that the paper comes up centered and straight...

Страница 28: ...the programme goes back to option 3 of the printer adjustment sub menu SUBMENU PRINTER 5 3 TEST Num 2 SPACE NEXT ENTER SELECT 5 4 Printer test n 3 This option starts the printer and prints simulated...

Страница 29: ...n 5 of the printer adjustment sub menu SUB MENU PRINTER 5 5 CUTTER TEST SPACE NEXT ENTER SELECT 5 6 Temperature of the printer This test shows the status of the printer head without printing Proceed a...

Страница 30: ...PACE NEXT ENTER SELECT 6 1 Programming the time Suppose that you wish to program 13 00 00 h Proceed as follows Key Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main...

Страница 31: ...he memory SUB MENU DATE TIME 6 2 DATE SPACE NEXT ENTER SELECT Note If during the time programming you need to cancel the process press the ESC key The keys and let you move the cursor through the disp...

Страница 32: ...left Moves the cursor to the right If you wish a double wide character press this key before the desired character CRTL C This combination does the same function as ENTER but it also centers the text...

Страница 33: ...if a card belongs to that premise or group of premises or not erase the data contained in the card charge new credits and list the recorded measurements and or the identification data of the card If...

Страница 34: ...the program goes back to option 1 of the smart card sub menu SMART CARD 8 1 CARD RESET SPACE NEXT ENTER SELECT 8 2 Programming the credit in the smart card This option is very similar to the previous...

Страница 35: ...ELECT 9 Voice Buzzer Test Option 9 of the main menu With this option we can check the correct operation of the audio system of the unit as well for the buzzer as for the voice messages The volume of t...

Страница 36: ...Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENTER SELECT 9 Select the voice and buzzer test SUB MENU VOICE BUZZ...

Страница 37: ...g is allright the menu goes to the previous screen In case of error the screen would be the one beside MODEM NOT RESPONDING PRESS ANY KEY TO END 10 1 Reset step by step This option is similar to the p...

Страница 38: ...case of error the screen would be the one beside MODEM NOT RESPONDING PRESS ANY KEY TO END 10 3 Central telephone With this option you indicate to the unit which is the telephone number of the Data Ce...

Страница 39: ...er of SMS to be deleted ERASE SMS INDICATE SMS NUMBER 01 ENTER If the SMS has been deleted correctly the message beside appears Press any key to continue or ENTER to end MESSAGE 1 DELETED Any Key The...

Страница 40: ...SMS SENDING SMS If the SMS has been sent succesfully the following message will appear SMS SENT TO DATA CENTER Any Key By pressing any key the system goes back to the selected option SUBMENU GSM SMS A...

Страница 41: ...Parameter 1 Parameter 1 indicates if the unit can send an SMS when switched on or when an error is detected Value Description Description 0 Does NOT send an SMS when unit is switched on 1 Sends SMS wh...

Страница 42: ...tion 0 Will never send an SMS for this reason 200 Sends an SMS with all data of status and collection and clears the collection counters when collection is 200 your currency etc Parameter 6 The unit w...

Страница 43: ...essary to inflate over 300 mmHg to measure the blood pressure 2 It was not possible to inflate the cuff Possible puncture in the cuff 3 Deflated up to 60mmHg without detecting impulsions 4 Before havi...

Отзывы: