Keito K7
Technical and maintenance
Manual
INT.022EN.02
Revision “I”
24/09/2010
Page 24 of 43
ENTER
When pressing ENTER, the systems asks you to place the 390
ohms resistor between the connectors.
CONNECT RESISTOR OF
390 ohms AND PRESS A
KEY OR ESC TO CANCEL
ENTER
With the resitor in place, press ENTER. The first step of the
calibration will start.
CALIBRATING
At the end of this step, the system asks you to place the 750 ohms
resistor between the connectors
CONNECT RESISTOR OF
750 ohms AND PRESS A
KEY OR ESC TO CANCEL
ENTER
With the resitor in place, press ENTER. The second step of the
calibration will start.
CALIBRATING
At the end of the process, the systems goes back to option
SUB-MENU FAT/PULSE:
4.1-CALIBRATION
SPACE=NEXT
ENTER=SELECT
4.2 - Heart rate analysis
To analyze the heart rate, the unit uses the same 4 electrodes used for the body fat estimation. In this case, the electrodes
record the electrical activity of the heart and from these signals the heart rate is measured. At the end of the test, the unit
prints a ticket with the value of the pulse rate detected and a graphic of the signal recorded.
Proceed as follows:
Key
Function to execute or select
Terminal Display
'Z' or 'ESC' Goes to option 0 of the main menu.
MAIN MENU:
0-SYSTEM ADJUSTMENTS
SPACE=NEXT
ENTER=SELECT
4
Select the body fat and heart rate adjustments.
SUB-MENU FAT/PULSE:
4.0-IMPEDANCE TEST
SPACE=NEXT
ENTER=SELECT
2
When pressing 2, the process starts. The user must grab the side
contacts or you have to connect the heart rate simulator. In this
case, connect the
RED
crocodiles on each of the two
upper
contacts and the
BLACK
crocodile on any of the
lower
contacts.
Press the RED button to cancel the process.
SUB-MENU FAT/PULSE:
4.2-PULSE TEST
SPACE=NEXT
ENTER=SELECT
At the end of the test, a ticket with the results of the test, similar to
the one shown below is printed.
SUB-MENU FAT/PULSE:
4.2- PULSE TEST
SPACE=NEXT
ENTER=SELECT