DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Section 3: Adjustment
DMM6500-905-01 Rev. E June 2021
3-13
cal_diode_100uA_zero_rear
cal_diode_10uA_zero_rear
Front-terminal adjustment with open circuit inputs
The following procedure provides instructions for completing a front-terminal adjustment with open
circuit inputs. The following section provides a command parameter table to complete the adjustment.
Prepare your DMM6500 for a front-terminal adjustment with open
circuit inputs
To prepare the DMM6500 for a front-terminal adjustment with open circuit inputs:
1. Set the TERMINALS switch to
FRONT
.
2. Remove all connections from the front terminals of your DMM6500.