DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Section 3: Adjustment
DMM6500-905-01 Rev. E June 2021
3-23
Command parameters
Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
stimulus value*
Reference
DMM dc
current range
cal_TS6
n/a
n/a
n/a
n/a
10 µA range (+) full scale nominal
cal_DCI_10uA_fs_pos
dc current 0.00022
0.00001
8508A or 8588A
Reading
10 µA
cal_DCI_DIGI_10uA_fs_pos
dc current 0.00022
0.00001
8508A or 8588A
Reading
10 µA
10 µA range (-) full scale nominal
cal_DCI_10uA_fs_neg
dc current 0.00022
-0.00001
8508A or 8588A
Reading
10 µA
cal_DCI_DIGI_10uA_fs_neg
dc current 0.00022
-0.00001
8508A or 8588A
Reading
10 µA
100 µA range (+) full scale nominal
cal_DCI_100uA_fs_pos
dc current 0.00022
0.0001
8508A or 8588A
Reading
100 µA
cal_DCI_DIGI_100uA_fs_po
s
dc current 0.00022
0.0001
8508A or 8588A
Reading
100 µA
100 µA range (-) full scale nominal
cal_DCI_100uA_fs_neg
dc current 0.00022
-0.0001
8508A or 8588A
Reading
100 µA
cal_DCI_DIGI_100uA_fs_ne
g
dc current 0.00022
-0.0001
8508A or 8588A
Reading
100 µA
*Characterized step measured by the Fluke 8508A or 8588A
Command parameters
Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
stimulus value*
Reference
DMM dc
current range
1 mA range (+) full scale nominal
cal_DCI_1mA_fs_pos
dc current 0.0022
0.001
8508A or 8588A
Reading
1 mA
cal_DCI_DIGI_1mA_fs_pos
dc current 0.0022
0.001
8508A or 8588A
Reading
1 mA
1 mA range (-) full scale nominal
cal_DCI_1mA_fs_neg
dc current 0.0022
-0.001
8508A or 8588A
Reading
1 mA
cal_DCI_DIGI_1mA_fs_neg
dc current 0.0022
-0.001
8508A or 8588A
Reading
1 mA
10 mA range (+) full scale nominal
cal_DCI_10mA_fs_pos
dc current 0.022
0.01
8508A or 8588A
Reading
10 mA
cal_DCI_DIGI_10mA_fs_pos
dc current 0.022
0.01
8508A or 8588A
Reading
10 mA