4-7
cal.password
This attribute sets the password that you send when you unlock calibration.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (W)
Yes
Not applicable
Nonvolatile memory
KI000CAL
Usage
cal.password = "
password"
password
A string that contains the password to unlock calibration; maximum of 10 characters
Details
This command can only be sent when calibration is unlocked.
The password is not saved until calibration is saved with
cal.save()
.
Be sure to record the password; there is no command to retrieve the password once it is set.
Example
cal.unlock("KI000CAL")
cal.password = "XYZCorp"
cal.lock()
cal.unlock("XYZCorp")
To change the default calibration password, unlock
the calibration with the default password.
Sets the password to
XYZCorp
.
Lock calibration. Use the password
XYZCorp
for
subsequent unlocks.
Also see
cal.save()
(on page 4-8)
cal.unlock()
(on page 4-9)