Section 3: Adjustment
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
3-14
DMM6500-905-01 Rev. E June 2021
Command parameters for a front-terminal adjustment with open circuit
inputs
When calibrating your DMM6500 for a front-terminal adjustment with open circuit inputs, use the
following command parameters.
Send each command parameter twice. First, send the parameter using the
setup
command. Second,
send the parameter using the
execute
command.
The example code below uses the first line of code from the table below. Send all command
parameters using this command.
cal.adjust.step.setup("cal_DCI_3A_zero_front")
cal.adjust.step.execute("cal_DCI_3A_zero_front")
Command parameters
cal_DCI_3A_zero_front
cal_DCI_1A_zero_front
cal_DCI_100mA_zero_front
cal_DCI_10mA_zero_front
cal_DCI_1mA_zero_front
cal_DCI_100uA_zero_front
cal_DCI_10uA_zero_front
cal_DCI_DIGI_3A_zero
cal_DCI_DIGI_1A_zero
cal_DCI_DIGI_100mA_zero
cal_DCI_DIGI_10mA_zero
cal_DCI_DIGI_1mA_zero
cal_DCI_DIGI_100uA_zero
cal_ACI_1A_zero
cal_ACI_3A_zero
cal_4W_HiOhm_7Vref_open
cal_CAP_zero
Rear-terminal adjustment with open circuit inputs
The following procedure provides instructions for completing a rear-terminal adjustment with open
circuit inputs. The following section provides a command parameter table to complete the adjustment.