background image

DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual 

Section 2: Performance verification 

 

DMM6500-905-01 Rev. E June 2021 

2-51 

 

AC current 10 A verification 

Verify that the displayed readings are within specified limits. 

Use the values in the tables following the steps below to verify the performance of the DMM6500. 

Actual values depend on the published specifications (see 

Example reading limit calculation

 (on page 

2-4)). 

To verify the ac current accuracy: 

1.  Set the TERMINALS switch to 

REAR

2.  Press the 

FUNCTION

 key, select the 

Measure Functions

 tab, and select 

AC Current

3.  Press the 

HOME

 key. 

4.  Set the range to 

10A

5.  Press the 

MENU

 key. 

6.  Under Measure, select 

Settings

7.  Make sure that detector bandwidth is set to 

30 Hz

AC current is specified for the detector bandwidth setting of 3 Hz. 3 Hz measures accurately for input 

signals from 3 

Hz to 10 kHz, with reading rates of ≈ 0.5 readings/s. To improve verification 

throughput to ≈ 3.3 readings/s, set detector bandwidth to 30 Hz for frequencies of 30 Hz to 10 kHz. 

To verify frequencies 1 kHz and higher, set the detector bandwidth to 300 Hz f

or faster ≈ 55 

readings/s throughput. 

Содержание DMM6500

Страница 1: ...6 Digit Multimeter with Scanning Calibration and Adjustment Manual DMM6500...

Страница 2: ...names are trademarks or registered trademarks of Keithley Instruments LLC Other brand names are trademarks or registered trademarks of their respective holders The Lua 5 0 software and associated docu...

Страница 3: ...d to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exerc...

Страница 4: ...lains hazards that could damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or m...

Страница 5: ......

Страница 6: ......

Страница 7: ...not defined AC voltage verification 2 10 Digitize dc voltage verification 2 14 Frequency verification 2 17 Simulated thermocouple Type J temperature verification 2 18 Simulated RTD temperature verifi...

Страница 8: ...13 Rear terminal adjustment with open circuit inputs 3 14 Resistance adjustment 3 15 DC voltage adjustment 3 18 DC current adjustment 3 20 DC current 10 A range adjustment 3 24 AC voltage adjustment 3...

Страница 9: ...calibrate and adjust your DMM6500 In this manual calibration refers to the process of verifying that the accuracy of the instrument is within its one year accuracy specifications Also adjustment refer...

Страница 10: ...fraction of the price of a repair Extended warranties are available on new and existing products Contact your local Keithley Instruments office sales partner or distributor for details Contact informa...

Страница 11: ...ent specifications You can use these verification procedures to Make sure that the instrument was not damaged during shipment Verify that the instrument meets factory specifications Determine if adjus...

Страница 12: ...ne voltage Using the proper test equipment Using the specified output signal and reading limits Environmental conditions Conduct the calibration verification procedures in a test environment with An a...

Страница 13: ...MM6500 specifications In this manual the Model 8610 shorting plug is shown in the figures However you can use either the Model 8610 or the Model 8620 shorting plug Manufacturer Model Description Used...

Страница 14: ...dc range using a 10 V input value Using the DMM6500 one year accuracy specification for 10 V dc of 25 ppm of reading 5 ppm of range the calculated limits are Reading limits 10 V 10 V 25 ppm 10 V 5 ppm...

Страница 15: ...ication on page 2 10 Digitize voltage verification on page 2 14 Frequency verification on page 2 17 Simulated thermocouple type J temperature verification on page 2 18 Simulated RTD temperature verifi...

Страница 16: ...ng equipment The front and rear terminals of the instrument are rated for connection to circuits rated Measurement Category II up to 300 V as described in International Electrotechnical Commission IEC...

Страница 17: ...es in the tables following the steps below to verify the performance of the DMM6500 Actual values depend on the published specifications see Example reading limit calculation on page 2 4 Use shielded...

Страница 18: ...ale voltages and allow for proper settling 14 Select each range on the DMM6500 allow for proper settling and verify the ranges according to the following tables Verify the dc voltage 100 mV range Desc...

Страница 19: ...Some of these procedures may expose you to hazardous voltages that if contacted could cause personal injury or death Use appropriate safety precautions when working with hazardous voltages Descriptio...

Страница 20: ...tables following the steps below to verify the performance of the DMM6500 Actual values depend on the published specifications see Example reading limit calculation on page 2 4 The maximum input volt...

Страница 21: ...ngs 6 Make sure that detector bandwidth is set to 30 Hz AC voltage is specified for the detector bandwidth setting of 3 Hz Three Hz measures accurately for input signals from 3 Hz to 300 kHz with read...

Страница 22: ...0017 1 100 kHz 9932 1 0068 Verify the ac voltage 10 V range Nominal value Frequency Lower limit Upper limit 10 20 Hz 9 991 10 009 10 1 kHz 9 991 10 009 10 50 kHz 9 983 10 017 10 100 kHz 9 932 10 068...

Страница 23: ...acy verification 750 V range 4 On the DMM6500 press the FUNCTION key and select AC voltage 5 On the home screen select the button next to Range and select 750 V 6 Press the MENU key 7 Under Measure se...

Страница 24: ...terminals on the front panel of the DMM6500 Verify that the displayed readings are within specified limits Use the values in the tables following the steps below to verify the performance of the DMM6...

Страница 25: ...select the button next to Range and select 100 mV 4 Press the MENU key 5 Under Measure select Settings 6 Set the Sample Rate to 1000 7 Set the Aperture to Auto 8 Set the Count to 100 9 Set the calibr...

Страница 26: ...49975 0 50025 Half scale 0 5 0 50025 0 49975 Full scale 1 1 0004 0 9996 Verify the digitize voltage 10 V range Description Nominal value Lower limit Upper limit Full scale 10 9 996 10 004 Half scale 5...

Страница 27: ...depend on the published specifications see Example reading limit calculation on page 2 4 1 Connect the Keithley Instruments Model 3390 function generator to the DMM6500 INPUT HI and LO terminals as s...

Страница 28: ...calibrator to the terminals on the front panel of the DMM6500 Verify that the displayed readings are within specified limits Thermocouple accuracy is verified by using a dc voltage calibrator to outpu...

Страница 29: ...erformance verification DMM6500 905 01 Rev E June 2021 2 19 To verify the simulated thermocouple type J temperature 1 Connect the DMM6500 HI and LO INPUT terminals to the calibrator HI and LO terminal...

Страница 30: ...to MENU Measure Calculations 15 Press the DMM6500 FUNCTION key and select Temperature 16 Press the MENU key 17 Under Measure select Settings 18 On the Measure Settings screen set the following values...

Страница 31: ...ar Van Dusen equation To verify RTD temperature accuracy you will Apply accurate resistance from the calibrator to the terminals on the front panel of the DMM6500 Verify that the displayed readings ar...

Страница 32: ...temperature in C A alpha 1 delta 100 B 1 alpha delta 1E 4 The alpha and delta values are listed in the following table RTD parameters for equations The RTD parameters for the Callendar Van Dusen equat...

Страница 33: ...values in the tables following the steps below to verify the performance of the DMM6500 Actual values depend on the published specifications see Example reading limit calculation on page 2 4 To verify...

Страница 34: ...0 25 C per 0 1 of HI LO lead resistance mismatch Figure 8 Connections for 3 wire RTD accuracy verification 3 On the DMM6500 press the FUNCTION key and select Temperature 4 Press the MENU key 5 Under...

Страница 35: ...value 19 18 999520 198 8900 0 0259 0 3241 100 99 99707 0 0075 0 0235 0 2932 190 189 99234 238 6775 0 0218 0 2725 Resistance verification Use the following information to verify the performance of the...

Страница 36: ...t Compensation On 7 Verify that Open Lead Detector is Off 8 On the home screen select the button next to Range and select 1 9 Source the nominal zero and full scale resistance values for the 1 to 10 k...

Страница 37: ...ameter measure range expected value from 1 to 100000 Set dmm measure offsetcompensation enable dmm OCOMP_OFF and then set dmm measure range 1e6 to run without warnings Verify that the readings are wit...

Страница 38: ...is with Input HI and LO and Sense LO only Range Nominal calibrator values Actual calibrator value Lower limit Upper limit 100 K 100 100 0012 99 9928 100 0098 1 M 0 0 1 1 1 0 999966 0 99986 1 000072 10...

Страница 39: ...n the home screen select the button next to Range and select 100 M 5 Source the nominal full scale resistance values for the 100 M range as shown in the following table The values and limits in the fo...

Страница 40: ...gurations are adequate for most purposes but may not provide sufficient test uncertainty ratio TUR for some users You must evaluate the measurement uncertainties and ensure that they are adequate for...

Страница 41: ...6500 905 01 Rev E June 2021 2 31 To prepare the DMM6500 for dc current accuracy verification 1 Set up the DMM6500 for dc current and the range being tested Make sure relative offset is disabled 2 Conn...

Страница 42: ...from the calculated reference DMM current 9 Repeat steps 1 through 8 for all ranges 10 A through 100 mA Test limit calculation for 10 A to 100 mA ranges The following tables list nominal test current...

Страница 43: ...000500 10 0 00250 DC current 100 mA to 3 A range verification To verify dc current accuracy on the 100 mA to 3 A ranges you will Apply accurate current from the dc current calibrator directly to the...

Страница 44: ...00 025 Half scale 50 49 985 50 015 Half scale 50 50 015 49 985 Full scale 100 100 025 99 975 Verify dc current 1 A range Description Calibrator setpoint A Lower limit Upper limit Full scale 1 0 99955...

Страница 45: ...as shown in the following figure Figure 13 Connections for digitize dc current 10 A to 3 A range verification 2 Press the FUNCTION key 3 Select the Digitize Functions tab and select Digitize Current...

Страница 46: ...r setpoint A Lower limit Upper limit Full scale 0 999993 0 998993 1 000993 Half scale 0 499998 0 499348 0 500648 Half scale 0 500009 0 500659 0 499359 Full scale 1 000017 1 001017 0 999017 Verify digi...

Страница 47: ...A range Description Calibrator setpoint A Lower limit Upper limit Full scale 2 1 997 2 003 Half scale 1 5 1 49745 1 50255 Half scale 1 5 1 50255 1 49745 Full scale 2 2 003 1 997 The 3 A range full sc...

Страница 48: ...ished specifications see Example reading limit calculation on page 2 4 To verify ac current accuracy 1 On the DMM6500 press the FUNCTION key and select AC Current 2 Press the HOME key 3 Set the range...

Страница 49: ...ing make sure that the reading is within low and high limits Verify ac current 100 A range Description Verification point Lower limit Upper limit 100 A at 20 Hz 100 99 83 100 17 100 A at 1 kHz 100 99...

Страница 50: ...9 kHz 100 99 86 100 14 Verify ac current 1 A range Description Verification point Lower limit Upper limit 1 A at 40 Hz 1 000 0 9986 1 0014 1 A at 1 kHz 1 000 0 9986 1 0014 1 A at 4 9 kHz 1 000 0 9986...

Страница 51: ...1 Rev E June 2021 2 41 Capacitance verification To compensate for capacitance offset of the cable and 1 F thru 100 F decade box 1 Connect the DMM6500 shielded banana cable banana to dual BNC cable and...

Страница 52: ...TION key select the Measure Functions tab and select Capacitance 4 Press the MENU key 5 Under Measure select Settings 6 Set the Range to 1 nF 7 Press the MENU key 8 Select Calculations and select Rel...

Страница 53: ...verification DMM6500 905 01 Rev E June 2021 2 43 9 Connect the shielded banana cable to the 1 nF to 1 F Decade Capacitance Box as shown in the figure below Figure 16 Capacitance verification connecti...

Страница 54: ...100 nF range 10 9 86 10 14 100 100 nF range 100 99 5 100 5 10 1 F range 0 1 0 0986 0 1014 100 1 F range 1 0 995 1 005 10 10 F range 1 0 986 1 014 100 10 F range 10 9 95 10 05 10 100 F range 10 9 86 1...

Страница 55: ...Settings 5 Set the Offset Compensation to On 6 Connect the Model 8610 or 8620 4 wire short to the front panel as shown in the following figure 7 Allow to settle for 5 minutes Do not use relative offse...

Страница 56: ...4 Press the HOME key 5 Set the range to 1000 V DC voltage verification is done in descending range order starting with the 1000 V range and finishing on the 100 mV range 6 Verify that the 1000 V rang...

Страница 57: ...2 Press the FUNCTION key and select DC Current 3 Press the HOME key 4 Set the range to 10A 5 Connect the DMM6500 reference DMM calibrator and amplifier as shown in the following figure Ensure cabling...

Страница 58: ...ey Ensure that the front panel displays STANDBY 2 Verify the DMM6500 zero reading 3 Verify the dc current verification points and reference DMM readings from the following table Verify dc current 10 A...

Страница 59: ...ent 10 A range verification Ensure cabling is designed to conduct 10 A without significant voltage drop In general cables for the 10 A range steps should be made of heavy gauge wire and should be as s...

Страница 60: ...the DMM6500 zero reading for each range 3 On the calibrator select the OPR STBY key Make sure that the front panel displays OPERATE 4 Enable the calibrator Amplifier and select the BOOST key 5 On the...

Страница 61: ...acy 1 Set the TERMINALS switch to REAR 2 Press the FUNCTION key select the Measure Functions tab and select AC Current 3 Press the HOME key 4 Set the range to 10A 5 Press the MENU key 6 Under Measure...

Страница 62: ...A range steps should be made of heavy gauge wire and should be as short as practical Figure 20 Connections for the 10 A range 9 Enable the calibrator amplifier and select the BOOST key 10 On the cali...

Страница 63: ...ion and Adjustment Manual Section 2 Performance verification DMM6500 905 01 Rev E June 2021 2 53 Verify ac current 10 A range Description Verification point Lower limit Upper limit 40 Hz 10 9 954 10 0...

Страница 64: ......

Страница 65: ...ed for a period of 90 days 1 year or 2 years from adjustment Adjustment should be performed at one of these intervals depending on your specification requirements Performance is specified relative to...

Страница 66: ...ss than or equal to 40 percent unless otherwise noted No direct airflow on the input terminals Line power The DMM6500 requires a line voltage of 100 V to 240 V and a line frequency of 400 Hz 50 Hz or...

Страница 67: ...table summarizes the recommended calibration equipment Specified accuracy of all functions and ranges is dependent on the precision of reference signals used during the adjustment process To achieve s...

Страница 68: ...g adjustment steps using a calibrator reference source be sure the calibrator reference signal has fully settled before sending adjustment commands to the DMM6500 The front and rear terminals of the i...

Страница 69: ...nt to the rear panel terminals press the TERMINALS switch on the front panel of the instrument Select the TSP command set Calibration adjustment must be performed by remote control using LAN USB or op...

Страница 70: ...using a combination of the two Interactive Manually set up and connect each reference signal then send the appropriate calibration adjustment command Programmatic A computer performs setup of calibrat...

Страница 71: ...parameters you need For a complete adjustment run through these sections in order making the preparations and running all of the command parameters You will send each command parameter twice The first...

Страница 72: ...with a 4 wire short 1 Set the TERMINALS switch to FRONT 2 Install the Keithley Model 8610 or 8620 shorting plug on the front terminals of the DMM6500 as shown in the figure below The shorting plug te...

Страница 73: ...ing this command cal adjust step setup cal_DCV_100mV_zero_front cal adjust step execute cal_DCV_100mV_zero_front Command parameters cal_DCV_100mV_zero_front cal_DCV_DIGI_100mV_zero cal_DCV_1V_zero_fro...

Страница 74: ..._4W_HiOhm_zero_sense cal_diode_10mA_zero_front cal_diode_1mA_zero_front cal_diode_100uA_zero_front cal_diode_10uA_zero_front cal_ACV_1V_zero cal_ACV_10V_zero Rear terminal adjustment with a 4 wire sho...

Страница 75: ...ment 1 Set the TERMINALS switch to REAR 2 Install the Keithley Model 8610 or 8620 shorting plug on the rear terminals of the DMM6500 as shown in the figure below The shorting plug terminals must be co...

Страница 76: ...t step setup cal_DCV_100mV_zero_rear_setup cal adjust step execute cal_DCV_100mV_zero_rear_setup Command parameters cal_DCV_100mV_zero_rear cal_DCV_1V_zero_rear cal_DCV_10V_zero_rear cal_sense_10V_zer...

Страница 77: ...g procedure provides instructions for completing a front terminal adjustment with open circuit inputs The following section provides a command parameter table to complete the adjustment Prepare your D...

Страница 78: ...parameters using this command cal adjust step setup cal_DCI_3A_zero_front cal adjust step execute cal_DCI_3A_zero_front Command parameters cal_DCI_3A_zero_front cal_DCI_1A_zero_front cal_DCI_100mA_ze...

Страница 79: ...wice First send the parameter using the setup command Second send the parameter using the execute command The example code below uses the first line of code from the table below Send all command param...

Страница 80: ...DMM6500 for a front terminal adjustment 1 Set the TERMINALS switch to FRONT 2 Connect the DMM6500 to the calibrator as shown in the following figure Figure 23 Connection for a resistance accuracy adju...

Страница 81: ...stimulus value The example code below uses the first line of code from the table below Send all command parameters using this command cal adjust step setup cal_4W_10ohm_fs cal adjust step execute cal...

Страница 82: ...tor cal_4W_10kohm_fs 4 wire 10000 10000 actual from calibrator cal_4W_10kohm_OCOMP_fs 4 wire 10000 10000 actual from calibrator cal_source_100uA_1V 4 wire 100000 100000 actual from calibrator cal_4W_1...

Страница 83: ...Rev E June 2021 3 19 Prepare your DMM6500 for a dc voltage adjustment To prepare the DMM6500 for a dc voltage adjustment 1 Connect a cable between the calibrator and the DMM6500 as shown in the figure...

Страница 84: ...ue cal_DCV_DIGI_1kV_fs_pos dc voltage 1100 1000 cal_DCV_DIGI_1kV_fs_neg dc voltage 1100 1000 cal_DCV_100V_fs_pos dc voltage 220 100 cal_DCV_DIGI_100V_fs_pos dc voltage 220 100 cal_DCV_100V_fs_neg dc v...

Страница 85: ...ual Section 3 Adjustment DMM6500 905 01 Rev E June 2021 3 21 Prepare your DMM6500 for a dc current adjustment To prepare the DMM6500 for dc current adjustment Connect the Model DMM6500 to the calibrat...

Страница 86: ...ommand parameters Send each command parameter twice First send the parameter using the setup command Second send the parameter using the execute command along with the calibrator stimulus value The ex...

Страница 87: ...100uA_fs_po s dc current 0 00022 0 0001 8508A or 8588A Reading 100 A 100 A range full scale nominal cal_DCI_100uA_fs_neg dc current 0 00022 0 0001 8508A or 8588A Reading 100 A cal_DCI_DIGI_100uA_fs_ne...

Страница 88: ...A_fs_po s dc current 0 22 0 1 n a 20 A cal_DCI_100mA_fs_neg dc current 0 22 0 1 n a 20 A cal_DCI_DIGI_100mA_fs_ne g dc current 0 22 0 1 n a 20 A cal_TS5 dc current n a n a n a 20 A cal_dcI_1A_fs_pos d...

Страница 89: ...6500 905 01 Rev E June 2021 3 25 Prepare your DMM6500 for a dc current 10 A range adjustment To prepare the DMM6500 for dc current adjustment for the 10 A range Connect the DMM6500 to the calibrator a...

Страница 90: ...e first line of code from the table below Send all command parameters using this command cal adjust step setup cal_DCI_10A_fs_pos cal adjust step execute cal_DCI_10A_fs_pos Command parameters Calibrat...

Страница 91: ...LS switch to FRONT 2 Connect a cable between the calibrator and the DMM6500 as shown in the figure below Figure 27 Connection for an ac voltage adjustment 3 Allow the instrument and cables to settle f...

Страница 92: ...voltage 0 22 0 1 100 kHz cal_ACV_100mV_200kHz_fs ac voltage 0 22 0 1 200 kHz cal_ACV_100mV_300kHz_fs ac voltage 0 22 0 1 300 kHz cal_ACV_1V_1kHz_1pct ac voltage 0 022 0 01 1 kHz cal_ACV_1V_10Hz_fs ac...

Страница 93: ...c voltage 22 7 300 kHz cal_ACV_700V_1kHz_fs ac voltage 1100 750 1 kHz AC current adjustment The following procedure provides instructions for completing a front terminal adjustment for ac current The...

Страница 94: ...rator frequency cal_ACI_100uA_1kHz_tenth ac current 0 00022 0 00001 1 kHz cal_ACI_100uA_1kHz_fs ac current 0 00022 0 0001 1 kHz cal_ACI_1mA_1kHz_tenth ac current 0 00022 0 0001 1 kHz cal_ACI_1mA_1kHz_...

Страница 95: ...range For this adjustment you need a Fluke 5720 calibrator The following section provides a command parameter table to complete the adjustment For this adjustment you need a Fluke 5720 calibrator 10...

Страница 96: ...ters using this command cal adjust step setup cal_ACI_10A_400Hz_tenth cal adjust step execute cal_ACI_10A_400Hz_tenth Command parameters Calibrator function Calibrator range Calibrator value Calibrato...

Страница 97: ...he DMM6500 for a frequency adjustment 1 Connect the Keithley Instruments Model 3390 function generator to the DMM6500 INPUT HI and LO terminals as shown in the following figure 2 Use the BNC to banana...

Страница 98: ...command parameters using this command cal adjust step setup cal_FREQ_1kHz cal adjust step execute cal_FREQ_1kHz Command parameter Calibrator function Calibrator range Calibrator value Calibrator freq...

Страница 99: ...cal adjust step setup cal_DCV_DIGI_100V_zero cal adjust step execute cal_DCV_DIGI_100V_zero Use front terminal 4 wire short setup cal adjust step setup cal_DCV_1kV_zero_front cal adjust step execute c...

Страница 100: ...inal 4 wire short setup cal adjust step setup cal_4W_100kohm_zero_front cal adjust step execute cal_4W_100kohm_zero_front Use front terminal 4 wire short setup cal adjust step setup cal_2W_1Mohm_zero_...

Страница 101: ...adjust step setup cal_4W_1ohm_zero_rear cal adjust step execute cal_4W_1ohm_zero_rear Use rear terminal 4 wire short setup cal adjust step setup cal_2W_10ohm_zero_rear cal adjust step execute cal_2W_1...

Страница 102: ...t step setup cal_4W_HiOhm_zero_rear cal adjust step execute cal_4W_HiOhm_zero_rear Use rear terminal 4 wire short setup cal adjust step setup cal_diode_10mA_zero_rear cal adjust step execute cal_diode...

Страница 103: ...en circuit setup cal adjust step setup cal_ACI_3A_zero cal adjust step execute cal_ACI_3A_zero Use front terminal 4 wire open circuit setup cal adjust step setup cal_4W_HiOhm_7Vref_open cal adjust ste...

Страница 104: ...up cal adjust step setup cal_source_1mA_1V cal adjust step execute cal_source_1mA_1V 1kohm_value_read_from_calibrator Use resistance setup cal adjust step setup cal_4W_10kohm_fs cal adjust step execut...

Страница 105: ...al adjust step setup cal_DCV_DIGI_10V_fs_neg cal adjust step setup cal_DCV_DIGI_10V_fs_neg Use dc voltage setup cal adjust step setup cal_DCV_DIGI_1V_fs_pos cal adjust step setup cal_DCV_DIGI_1V_fs_po...

Страница 106: ..._from_ 8508A 8588A Use dc current setup cal adjust step setup cal_DCI_DIGI_10mA_fs_pos cal adjust step execute cal_DCI_DIGI_10mA_fs_pos value_read_from_ 8508A 8588A Use dc current setup cal adjust ste...

Страница 107: ...V_100mV_10Hz_fs cal adjust step execute cal_ACV_100mV_10Hz_fs Use ac voltage setup cal adjust step setup cal_ACV_100mV_1kHz_fs cal adjust step execute cal_ACV_100mV_1kHz_fs Use ac voltage setup cal ad...

Страница 108: ...cal_ACV_10V_300kHz_fs Use ac voltage setup cal adjust step setup cal_ACV_100V_10Hz_1pct cal adjust step execute cal_ACV_100V_10Hz_1pct Use ac voltage setup cal adjust step setup cal_ACV_100V_1kHz_1pct...

Страница 109: ...1kHz_tenth cal adjust step execute cal_ACI_10mA_1kHz_tenth Use ac current setup cal adjust step setup cal_ACI_10mA_10Hz_fs cal adjust step execute cal_ACI_10mA_10Hz_fs Use ac current setup cal adjust...

Страница 110: ...0A_5kHz_fs Use ac current 10 A setup cal adjust step setup cal_ACI_10A_400Hz_2A cal adjust step execute cal_ACI_10A_400Hz_2A Use ac current 10 A setup cal adjust step setup cal_ACI_10A_10Hz_2A cal adj...

Страница 111: ...fy date os time year 2018 month 1 day 9 The cal verify date command is used to record the date of the last verification that was done independently of DMM6500 adjustment Typically this date is set at...

Страница 112: ...prompts mode is not recommended for automated test programs If prompts mode is enabled in an automated test the control program must be constructed to expect and read each prompt that is sent after e...

Страница 113: ...4 1 TSP commands The TSP commands available for the instrument are listed in alphabetical order Introduction This section contains detailed information on the DMM6500 remote calibration commands Sect...

Страница 114: ...applicable Usage adjustments cal adjust count adjustments The number of adjustments Details You can use this command if calibration is locked or unlocked The adjust count is read only The count is aut...

Страница 115: ...adjustDate The date when the last adjustment occurred year Year must be more than 1970 month Month 1 to 12 day Day 1 to 31 hour Hour in 24 hour time format 0 to 23 minute Minute 0 to 59 second Second...

Страница 116: ...ved Default value Function Yes Usage cal adjust step setup stepname cal adjust step setup stepname value stepname The adjustment step to start value The value for this adjustment step If using the def...

Страница 117: ...aved Default value Function Yes Usage cal adjust step execute stepname cal adjust step execute stepname value stepname The adjustment step to start value Value for this adjustment step If using the de...

Страница 118: ...To perform calibration you must unlock calibration This command does not save calibration data Calibration data is lost if it is you do not save it before sending cal lock Use cal save to save the dat...

Страница 119: ...tains the password to unlock calibration maximum of 10 characters Details This command can only be sent when calibration is unlocked The password is not saved until calibration is saved with cal save...

Страница 120: ...o sets the adjustment date and increments the adjustment count Calibration constants are retained indefinitely once saved Calibration is temporary unless the changes are saved Calibration data is not...

Страница 121: ...bration Details Calibration data is locked during normal operation To perform calibration you must unlock calibration The default password is KI000CAL You can use cal password to change the default An...

Страница 122: ...last verification occurred year Year must be more than 1970 month Month 1 to 12 day Day 1 to 31 hour Hour in 24 hour time format 0 to 23 minute Minute 0 to 59 second Second 0 to 59 Details The date an...

Страница 123: ...1 4 11 Example 1 cal verify date os time year 2018 month 9 day 5 print cal verify date Set the verify calibration date to September 5 2018 Verify the date Example output Sep 5 2014 12 00 00 000 Exampl...

Отзывы: