Section 2: Performance verification
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
2-52
DMM6500-905-01 Rev. E June 2021
8. Connect the DMM6500 to the calibrator and amplifier as shown in the following figure.
Ensure cabling is designed to conduct 10 A without significant voltage drop. In general,
cables for the 10 A range steps should be made of heavy gauge wire and should be as short
as practical.
Figure 20: Connections for the 10 A range
9. Enable the calibrator amplifier and select the
BOOST
key.
10. On the calibrator, set the amplifier range lock to
ON
.
11. Source ac current for each of the frequencies listed in the following table.
12. On the calibrator, select
OPR/STBY
key. Make sure that the front panel displays
OPERATE
.
13. For each setting, allow the calibrator and amplifier to settle properly and verify the reading is
within low and high limits.