Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Section 3: Pulse Current Measurements
:DELay <NRf>
Specify trigger delay in seconds:
0.0–0.1 (pulse current measurements) or
0.0–5.0 (pulse current digitization).
0.0
:STEP <b>
Performs a series of measurements
(
Pulse current
step method
)
OFF
:UP <NRf>
<0-20> (max is for both up and down combined)
1
:DOWN <NRf>
<0-20> (max is for both up and down combined)
1
:TIME <NRf>
33.3
μ
sec–100msec
200
μ
sec
:TimeOUT <NRf>
TimeOUT (other than the first): 2msec–200msec
2 ms
:INITial <NRf>
First TimeOUT step: 10msec–60secs
2 sec
:DELay <NRf>
0msec–100msec (in 10msec steps)
0
:SEQuence
<NRf, NRf>
Define up to 19 delay sequence values using the
index and the time value (e.g. "1,1e-3" sets the first delay to
1ms). Ranges: 1-19 (parameter 1) and 0 to 100e-3
(parameter 2).
0
:LENgth <NRf>
Define the delay sequence length (0 to 19). Specify
zero to always use the
PCUR:STEP:DELay
value.
0
:SKIP <b>
Specify whether (1) or not (0) to skip sequential
triggers.
1=Only requires 1 trigger
0 = Requires multiple triggers
0
:TLEVx <NRf>
Set trigger level for each TLEV step where x equals
1–20 (the maximum settings are based on the selected
current range. Therefore, the current range must be
selected before setting the step trigger level values)
0.0
:FAST
Enable or disable pulse current fast readings.
OFF
:SEARch
Enable or disable pulse current search.
ON
:DETect
Enable or disable pulse current detection mode.
OFF
:TimeOUT
Specify length of timeout: 5ms - 32s incrementing in
1ms.
1 (sec)
SENSe2
SENSe subsystem for charger channel (#2):
:FUNCtion “PCURrent”
Select pulse current measurement function.
VOLT
:PCURrent
Pulse current configuration:
:AVERage <NRf>
Specify average count:
1 to 100 (pulse current measurements), or
1 to 5000 (pulse current digitization).
1
:MODE <name>
Select measurement mode; HIGH, LOW or AVERage.
HIGH
:TIME
Set integration times:
:AUTO
Integration times set automatically.
:HIGH <NRf>
Specify integration time (in sec) for high pulse
measurements; 33.33e-6 to 0.8333.
3.333e-5
Table 3-3
SCPI commands - pulse current measurements (cont.)
Command
Description
Default
Содержание 2308
Страница 2: ......
Страница 12: ...This page left blank intentionally...
Страница 32: ...This page left blank intentionally...
Страница 106: ...This page left blank intentionally...
Страница 140: ...This page left blank intentionally...
Страница 208: ...This page left blank intentionally...
Страница 209: ...Appendix A Error and Status Messages In this Appendix Error and status message definitions A 2 Topic Page...
Страница 220: ...This page left blank intentionally...