Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Section 2: Basic Power Supply Operation
4. Press
▲
,
▼
, ,
or
keys to scroll to the desired output impedance value.
5. Press
ENTER
to save your settings and return to the main menu.
SCPI programming - outputting voltage and current
The commands to output voltage and current are summarized in
Table 2-3
(a listing following the
table contains specific command notes). The programming example (“Outputting and reading back
V and I”) located at the end of this section demonstrates how to use these commands.
NOTE
Setting changes can be canceled by pressing the
MENU
key.
NOTE
Brackets [ ] indicate optional (and default) command parameters
.
Table 2-3
SCPI command summary - outputting voltage and current
Commands
Description
Default
SENSe[1]
:CURRent
:RANGe
[:UPPer] <n>
:AUTO <b>
SENSe[1] subsystem for battery channel (#1):
Current function:
Set current measurement range:
Specify expected current in amps: 0 to 5.
Enable or disable auto range.
5.0
OFF
SENSe2
:CURRent
:RANGe
[:UPPer] <n>
:AUTO <b>
SENSe2 subsystem for charger channel (#2):
Current function:
Set current measurement range:
Specify expected current in amps: 0 to 5.
Enable or disable auto range.
5.0
OFF
[SOURce[1]]
:VOLTage <n>
:PROTection <NRf>
:STATe? <b>
:CLAMp <b>
:CURRent <n>
:TYPe <name>
:STATe?
[SOURce1] subsystem for battery channel (#1):
Set voltage amplitude in volts: 0 to 15 (1mV resolution).
Sets VPT (voltage protection) range (0–8V).
Query state of VPT - no associated command.
Sets VPT clamp mode ON or OFF.
Set current limit value in amps: 0.006 to 5 (100µA res)
Select current limit type: LIMit or TRIP.
Query state of current limit - no associated command.
0.0
8V
OFF
0.25
LIM
SOURce2
:VOLTage <n>
:PROTection <NRf>
:STATe? <b>
:CLAMp <b>
:CURRent <n>
:TYPe <name>
:STATe?
SOURce2 subsystem for charger channel (2):
Set voltage amplitude in volts: 0 to 15 (1 mV resolution).
Sets VPT range (0–8V).
Query state of VPT - no associated command.
Sets VPT clamp mode ON or OFF.
Set current limit value in amps: 0.006 to 5 (100 µA res).
Select current limit type: LIMit or TRIP.
Query state of current limit - no associated command.
0.0
8V
OFF
0.25
LIM
OUTPut[1]
[:STATe] <b>
:BANDwidth <name>
:IMPedance <NRf>
OUTPut [1] subsystem for battery channel (#1):
Turn the power supply output ON or OFF.
Specifies output bandwidth (HIGH or LOW).
Specifies output impedance (0–1
Ω
in 10 m
Ω
steps).
OFF
LOW
0
OUTPut2
[:STATe] <b>
:BANDwidth <name>
OUTPut2 subsystem for charger channel (#2):
Turn the power supply output ON or OFF.
Specifies output bandwidth (HIGH or LOW).
OFF
LOW
▲
▲
Содержание 2308
Страница 2: ......
Страница 12: ...This page left blank intentionally...
Страница 32: ...This page left blank intentionally...
Страница 106: ...This page left blank intentionally...
Страница 140: ...This page left blank intentionally...
Страница 208: ...This page left blank intentionally...
Страница 209: ...Appendix A Error and Status Messages In this Appendix Error and status message definitions A 2 Topic Page...
Страница 220: ...This page left blank intentionally...