3-14
Return to
Section Topics
2308-900-01 Rev. A / July 2008
Section 3: Pulse Current Measurements
Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Command notes (pulse current measurements)
SENSe[1]:FUNCtion ‘PCURrent’
Applies to battery channel (#1)
SENSe2:FUNCtion ‘PCURrent’
Applies to charger channel (#2)
This parameter name can also be enclosed in single quotes (as shown above).
SENSe[1]:PCURrent:AVERage <NRf>
Applies to battery channel (#1)
SENSe2:PCURrent:AVERage <NRf>
Applies to charger channel (#2)
1. When requesting a single reading (
FETch?
,
READ?
, or
MEASure?
), average count
specifies the number of pulse current measurement conversions to average for the reading.
For example, with the average count set to 10,
READ?
will trigger 10 pulse current
measurement conversions and return (and display) the average of those 10 battery channel
conversions (charger channel command similar).
2. When requesting an array of readings (
FETCh:ARRay?
,
READ:ARRay?
or
MEASure:ARRay?), average count specifies the number of pulse current measurements to
place in an array. For example, with the average count set to 10,
READ:ARRay?
will trigger
and return 10 battery channel readings (charger channel command similar).
3. For pulse current digitization, use an array reading command (such as
READ:ARRay?
) to
return the digitized readings.
4. Signal-oriented measurement commands (such as
READ?
) are covered in
Section 9
.
:LOW <NRf>
Specify integration time (in sec) for low pulse
measurements; 33.33e-6 to 0.8333.
3.333e-5
:AVERage <NRf>
Specify integration time (in sec) for average pulse
measurements; 33.33e-6 to 0.8333.
3.333e-5
:DIGitize <NRf>
Specify integration time (in sec) for digitizing or burst
measurements 33.33e-6 to 0.8333.
3.333e-5
:SYNChronize
Pulse detection triggering:
[:STATe]
Send ON to select pulse current measurements or
OFF to select pulse current digitization.
ON
:TLEVel <NRf>
Set trigger level in amps: 0.0–5.0
0.0
:DELay <NRf>
Specify trigger delay in seconds:
0.0–0.1 (pulse current measurements) or
0.0–5.0 (pulse current digitization).
0.0
:FAST
Enable or disable pulse current fast readings.
OFF
:SEARch
Enable or disable pulse current search.
ON
:DETect
Enable or disable pulse current detection mode.
OFF
:TimeOUT
Specify length of timeout: 5ms - 32s incrementing in
1ms.
1 (sec)
READ[1]?
Trigger and return one reading for battery channel (#1).
READ[1]:ARRay?
Trigger an array of readings and return them for battery
channel (#1).
READ2?
Trigger and return one reading for charger channel (#2).
READ2:ARRay?
Trigger an array of readings and return them for charger
channel (#2).
Table 3-3
SCPI commands - pulse current measurements (cont.)
Command
Description
Default
Содержание 2308
Страница 2: ......
Страница 12: ...This page left blank intentionally...
Страница 32: ...This page left blank intentionally...
Страница 106: ...This page left blank intentionally...
Страница 140: ...This page left blank intentionally...
Страница 208: ...This page left blank intentionally...
Страница 209: ...Appendix A Error and Status Messages In this Appendix Error and status message definitions A 2 Topic Page...
Страница 220: ...This page left blank intentionally...