7-16
Return to
Section Topics
2308-900-01 Rev. A / July 2008
Section 7: Status Structure
Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Programming example - read error queue
STAT:QUE:ENAB (+000:+900)
‘ Enable all Keithley defined messages (disable all SCPI
defined messages).
STAT:QUE?
‘ Return oldest message.
Table 7-7
SCPI Commands - error queue
Command
Description
Default
STATus
:QUEue
[:NEXT]?
:ENABle <list>
:ENABle?
:DISable <list>
:DISable?
:CLEar
STATus subsystem:
Read error queue:
Read and clear oldest error/status message.
Specify error and status messages for error
queue.
Read the enabled messages.
Specify messages not to be placed in queue.
Read the disabled messages.
Clear messages from error queue.
(Note 1)
(Note 2)
(Note 2)
SYSTem
:ERRor?
:CLEar
SYSTem subsystem:
Read error queue:
Clear messages from error queue.
(Note 1)
Notes:
1. Power-up and *CLS empties the error queue.
STATus:PRESet
has no effect.
2. Power-up enables error messages and disables status messages. *CLS and STATus:PRESet
have no effect.
Содержание 2308
Страница 2: ......
Страница 12: ...This page left blank intentionally...
Страница 32: ...This page left blank intentionally...
Страница 106: ...This page left blank intentionally...
Страница 140: ...This page left blank intentionally...
Страница 208: ...This page left blank intentionally...
Страница 209: ...Appendix A Error and Status Messages In this Appendix Error and status message definitions A 2 Topic Page...
Страница 220: ...This page left blank intentionally...