Section 12: Performance Verification
Digital voltmeter input accuracy
Follow the steps below to verify the digital voltmeter input accuracy is within specified limits. The
test involves setting the voltage applied to the DVM input to accurate values and then verifying that
the digital voltmeter input readings are within required limits.
1. With the power supply’s output off, connect the OUTPUT #2 DVM IN, SOURCE, and
SENSE terminals to the digital multimeter, as shown in
Figure 12-6
. Be sure to observe
proper polarity (DVM IN +, and, SENSE + to DMM INPUT HI; also DVM IN -,
SOURCE -, and SENSE - to DMM INPUT LO).
Figure 12-6
Charger channel connections for DVM accuracy verification
2. Select the DMM DC volts function, and enable auto-ranging.
3. Press
the
or keys to select Channel
#2
.
Enable the DVM input.
4. Turn on the source output.
5. Verify digital voltmeter input accuracy for each of the voltages listed in
Table 12-9
. For each
test point:
6 V
2.0000 mA
1.9950 to 2.0050 mA
9 V
3.0000 mA
2.9930 to 3.0070 mA
12 V
4.0000 mA
3.9910 to 4.0090 mA
14.7 V
4.9000 mA
4.8892 to 4.9108 mA
Table 12-8
5 mA range current readback accuracy limits
*
Nominal output
Voltage
Model 2308
output current
**
Current readback limits
(2 Years, 18°C-28°C)
*Check the complete Model 2308 Specifications at
www.keithley.com
for updated
current readback accuracy limits before verification.
** As measured by digital multimeter voltage reading and 3 k
Ω
resistance value.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-120VAC, 200-240VAC
50, 60 HZ 165VA MAX
LINE FUSE
SLOWBLOW
2.0A, 250V
IEEE-488
(ENTER IEEE ADDRESS
FROM FRONT PANEL MENU)
REMOTE
DISPLAY
OPTION
+
+
+ _
_ _
SOURCE
SENSE
SOURCE
OUTPUT #1
ISOLATION FROM EARTH: 22 VOLTS MAX MAX.
RELAY
CONTROL
24VDC MAX
DVM IN
-30 VDC MAX
ANALOG
OUT
mA
(5/50)
A
(0.5/5)
+
+
+ _
_ _
_
+
SOURCE
SENSE
DVM IN
SOURCE
OUTPUT #2
NEXT
PREV
POWER
DISPLAY
2001
MULTIMETER
DCV
ACV
DCI
ACI
ý2
ý4
FREQ
TEMP
REL
TRIG
INFO LOCAL
EXITENT
ER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
ý 4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
Model 2001 DMM
Model 2308
Input LO
Input HI
Source -
DVM IN -
DVM IN +
Sense +
Sense -
▲
▲
Содержание 2308
Страница 2: ......
Страница 12: ...This page left blank intentionally...
Страница 32: ...This page left blank intentionally...
Страница 106: ...This page left blank intentionally...
Страница 140: ...This page left blank intentionally...
Страница 208: ...This page left blank intentionally...
Страница 209: ...Appendix A Error and Status Messages In this Appendix Error and status message definitions A 2 Topic Page...
Страница 220: ...This page left blank intentionally...