2-18
Return to
Section Topics
2308-900-01 Rev. A / July 2008
Section 2: Basic Power Supply Operation
Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Command notes (measure V and I, and DVM input)
SENSe[1]:FUNCtion <name>
Applies to battery channel (#1)
SENSe2:FUNCtion <name>
Applies to charger channel (#2)
The parameter name can instead be enclosed in single quotes (for example: ‘CURRent’).
With “DVMeter” selected, the instrument measures the voltage applied to the input of the digital
voltmeter (DVM) on the charger channel (#2) only.
The “
PCURrent
” and “
LINTegration
” parameters for
:FUNCtion
(which are not listed in
Table 2-
4
) select the pulse current and long integration measurement modes. These measurement modes
are covered in
Section 3
and
Section 4
, respectively.
SENSe[1]:AVERage <NRf>
Applies to battery channel (#1)
SENSe2:AVERage <NRf>
Applies to charger channel (#2)
• When requesting a single reading (
FETch?
,
READ?
, or
MEASure?
), average count
specifies the number of measurement conversions to average for the reading. For example,
with the average count set to 10,
READ?
will trigger 10 measurement conversions and
return (and display) the average of those 10 conversions for the battery channel. When
requesting an array of readings (
FETCh:ARRay?
,
READ:ARRay?
or
MEASure:ARRay?
),
average count specifies the number of measurements to place in an array. For example,
with the average count set to 10,
READ:ARRay?
will trigger and return 10 battery channel
readings (charger channel command similar).
• Signal-oriented measurement commands (for example:
READ?
) are covered in
Section 9
.
Independent voltage measurements (DVM)
The power supply has an independent digital voltmeter (DVM) on the charger channel (# 2) only
that can measure up to +30VDC and down to -5VDC. Connections for the DVM are shown in
Figure 2-1
.
DVM input display mode
The DVM input display mode must be selected in order to measure voltage applied to DVM input
of the power supply.
This display mode is selected as follows:
1. Press
the
DISPLAY
key to access the display menu.
2. Press
the
or keys to scroll to
DISPLAY TYPE #2
(DVM is a Channel 2 only feature).
3. Press
the
▲
or
▼
keys to scroll to
DVM INPUT
.
4. Press
ENTER
to select.
NOTE
To display measured readings if the instrument is in the settings mode, press the
SET
key until the blinking stops (the measured readings can then be displayed).
To determine if the instrument is in the settings mode, check for a blinking cursor
in a digit of the voltage or current field (if present, the instrument is in the setting
mode).
NOTE
For details about display modes, see
Display modes
.
▲
▲
Содержание 2308
Страница 2: ......
Страница 12: ...This page left blank intentionally...
Страница 32: ...This page left blank intentionally...
Страница 106: ...This page left blank intentionally...
Страница 140: ...This page left blank intentionally...
Страница 208: ...This page left blank intentionally...
Страница 209: ...Appendix A Error and Status Messages In this Appendix Error and status message definitions A 2 Topic Page...
Страница 220: ...This page left blank intentionally...