4-12
Return to
Section Topics
2308-900-01 Rev. A / July 2008
Section 4: Long Integration Measurements
Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Command notes (long integration measurements)
SENSe[1]:FUNCtion ‘LINTegration’
Applies to battery channel (#1)
SENSe2:FUNCtion ‘LINTegration’
Applies to charger channel (#2 )
The parameter name can be enclosed in single or double quotes (single - shown above, double -
shown in
Table 4-2
).
SENSe[1]:LINTegration:SEARch <b>
Applies to battery channel (#1)
SENSe2:LINTegration:SEARch <b>
Applies to charger channel (#2 )
Refer to
Using FAST, SEARch, and DETect
for detailed usage information.
SENSe[1]:LINTegration:FAST <b>
Applies to battery channel (#1)
SENSe2:LINTegration:FAST <b>
Applies to charger channel (#2 )
Refer to
Using FAST, SEARch, and DETect
for detailed usage information.
SENSe[1]:LINTegration:DETect <b>
Applies to battery channel (#1)
SENSe2:LINTegration:DETect <b>
Applies to charger channel (#2 )
Refer to
Using FAST, SEARch, and DETect
for detailed usage information.
READ[1]?
Applies to battery channel (#1)
READ2?
Applies to charger channel (#2 )
After sending a trigger reading command to perform long integration measurements, do not
address the power supply to talk until all readings are completed. Details on
READ?
and the other
signal-oriented measurement commands are provided in
Section 9
.
Using FAST, SEARch, and DETect
Use the
FAST
command to have the quickest response for pulse current measurements triggered
by the measurement commands (for example,
MEASure
,
READ,
or *
TRG)
. Note that when you
use the
FAST
command, there is no indication that the instrument is correctly configured for
detecting pulses until a measurement command is sent. When the trigger level is set, use the
SEARch
command to retain quick responses to measurement commands and to ensure that the
instrument is configured correctly for detecting pulses. Use the
DETect
command to ensure that
the unit is configured correctly and can read the pulse before sending a measurement command.
:TEDGe <name>
Select trigger edge to initiate the
measurement:
RISING, FALLING, or NEITHER.
RISING
:TimeOUT <NRf>
Specify length of timeout: 1–63 (sec).
16
:SEARch <b>
Enable or disable pulse search.
ON
:FAST <b>
Enable or disable long integrations fast
readings.
OFF
:DETect <b>
Enable or disable pulse detection mode.
OFF
READ[1]?
Trigger and return one reading for battery channel
(#1).
READ2?
Trigger and return one reading for charger channel
(#2).
Table 4-2
SCPI commands - long integration measurements
Command
Description
Default
Содержание 2308
Страница 2: ......
Страница 12: ...This page left blank intentionally...
Страница 32: ...This page left blank intentionally...
Страница 106: ...This page left blank intentionally...
Страница 140: ...This page left blank intentionally...
Страница 208: ...This page left blank intentionally...
Страница 209: ...Appendix A Error and Status Messages In this Appendix Error and status message definitions A 2 Topic Page...
Страница 220: ...This page left blank intentionally...