2-2
Return to
Section Topics
2308-900-01 Rev. A / July 2008
Section 2: Basic Power Supply Operation
Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Test connections
Test connections to the power supply are made at the rear panel using a quick disconnect
connector (see rear panel in
Figure 1-1
for connector location). Use up to #14 AWG wire for the
screw terminals of the connector. Once the connector is wired up, plug it into the rear panel and
tighten the captive retaining screws.
Figure 2-1
and
Figure 2-2
show optimum remote 4-wire connections for device-under-test (DUT)
preferred connection (
Figure 2-1
) and fastest transient response connection (
Figure 2-2
).
WARNING
When installing a unit into a test system, make sure the external power
sources do not apply voltage to the power supply in excess of its
maximum limits (see complete Specifications at
www.keithley.com
). Failure
to do so could result in personal injury or death.
The source and measurement connections are provided with overvoltage
protection rated up to 500 V for 50 µs. Do not connect sources that
produce transient voltages greater than 500 V or the protection provided
by the equipment may be degraded.
Содержание 2308
Страница 2: ......
Страница 12: ...This page left blank intentionally...
Страница 32: ...This page left blank intentionally...
Страница 106: ...This page left blank intentionally...
Страница 140: ...This page left blank intentionally...
Страница 208: ...This page left blank intentionally...
Страница 209: ...Appendix A Error and Status Messages In this Appendix Error and status message definitions A 2 Topic Page...
Страница 220: ...This page left blank intentionally...