Eliminating disturbances
Note
If disturbances cannot be located or eliminated by using this troubleshooting
method, a technician authorised by KaVo must try and eliminate the problem.
Malfunction
Cause
Remedy
Treatment unit and oper‐
ating light do not function
The main switch of the
treatment unit is off
Turn the main switch of
the treatment unit on
The treatment light turns
off and on sporadically
The thermal fuse for pro‐
tecting the operating light
is tripped
The thermal fuse turns on
after the light cools. The
fan may not work, or the
circulation is interrupted
from dirt and dust
The operating light does
not work
The halogen lamp is de‐
fective
Exchange the halogen
lamp
(Mat. no. 0.740.4211)