▶ To start service mode, press and hold the "Decrease value" and "Increase val‐
ue" keys.
▶ Press the foot pedal.
Service mode U1 is shown on the display.
▶ Press the "Select level" button until "U13" is shown on the display.
▶ Press the "Decrease value" and "Increase value" keys to set the brightness.
The setting range is 0 through 50, and the standard value is 25.
▶ Measure the lamp voltage of the halogen lamp, change the correction value
with the service tester I
2
C if necessary, and set it to the voltage values descri‐
bed above.
▶ To quit the service mode, press and hold the "Decrease value" and "Increase
value" keys.
▶ Press the foot pedal.
This quits the service mode.
4.3.3
ESTETICA E80 to Ser. No.: 10100000
See also: TA treatment unit: Start, continue and end the service mode
▶ Press and hold down the "Timer menu" and "Patient menu" keys.
▶ Press "MEMOdent menu" key briefly.