▶ Loosen screws ③ on the disinfection container and remove the container ④.
▶ Ensure that the protective earth screw ⑥ and fastening screws ⑤ and ⑦ do
not reach into the light holder.
Note
When the mounting rod extends into the light holder, it can damage the paint.
Screw in the depth stop screw ⑧ completely.
▶ Slide the cover ① (Mat. no. 0.764.7643) onto the mounting pole ②.
▶ Insert the mounting pole into the lamp holder.
Positioning the mounting pole:
▪ It is approx. 1800 mm from the top edge to the floor.
▪ The borehole for the swivel arm brake points in the direction 6 o´ clock
▶ Tighten the earth protector screw ⑥ and fastening screws ⑤ and ⑦.
▶ Return the disinfection container ④ to its former position and fix with screws
③.
Instructions for use KAVOLUX 1410 C
4 Assembly | 4.2 Mounting the device