4.3.1
ESTETICA E50, ESTETICA E70 ab Ser. No.: 20100001 and
ESTETICA E80 from Ser. No.: 10100001
▶ Press and hold down the "Function level" key.
▶ Press the "Remote Control" button.
Service mode appears on the display. If the service mode was active already, this
terminates the service mode.
▶ Press the "Remote Control" button until "U10" is shown on the display.
Function key
Description
S1
Move cursor
S2
Reduce value
S3
Increase value
S6
Save value
▶ Set the voltage at the lamp to 12.5VAC with a normal voltmeter.
or
▶ Set the voltage at the lamp to 16.5VAC with an RMS normal voltmeter.
▶ After setting the correction value check the voltage at the light.
4.3.2
ESTETICA E70 to Ser. No.: 20100000
Note
The KaVo Service Tester (Mat. no. 1.003.2212) is required to make adjustments
in service mode.
▪ Service tester: Service tester I²C Mat. no. 1.003.2212
See also: TA treatment unit: Start, continue and end the service mode