Note
Always save the selected setting with enter button when leaving the menu. Other‐
wise, the motor will not start (safety circuit).
Examples of saving values of handpieces and contra-angle handpie‐
ces
Note
Default values for the respective transfer ratios can be selected with the Plus or
Minus buttons.
Indication
Transfer ra‐
tio
Combined
unit
Reducing
shank
Head
Speed
range
rpm
Maximum
torque
Ncm
Grinding,
cutting
1:2
3555
1:2
--
600-80.000 2,7
Grinding,
Drilling,
cutting
1:1
3,610 N
1:1
--
300-40.000 5,5
Grinding,
Drilling,
Thread,
Implant
12:1
3624 N
4:1
C 3
3:1
25-3.300
40
Grinding,
Drilling,
Thread,
Implant
16:1
3624 N
4:1
67 RIC
4:1
18-2.500
45
Grinding,
Drilling,
Thread,
Implant
27:1
C 09
9:1
C 3
3:1
11-1.500
55
Grinding,
Drilling,
Thread,
Implant
36:1
C 09
9:1
67 RIC
4:1
9-1.100
55
Note
The listed indications are only examples. To prevent unnecessary risk, observe the
guideline speeds given by the manufacturer of the rotating instruments. KaVo rec‐
ommends setting a speed range of 11-25/min or 11-35/min for screwing tasks.