PRG
▶ Press the program button for at least three seconds.
An acoustic signal confirms the entry.
5.4.6
Setting up delivery status
Note
All entries and changes made are deleted with this function.
Note
If the device was initialised, everything saved previously is deleted.
Therefore, for simplicity, everything that was saved should be noted down for easy
recovery.
PRG
▶ Turn device on while pressing the programme button.
After initialization, the device is ready for use in its condition at the time of de‐
livery.
Condition at the time of delivery: Programme
Step
Instrument
Speed (rpm) Torque
(Ncm)
Pump
Direction of
motor rota‐
tion
1
27:1
11-1500
55
4
->
2
27:1
11-50
40
4
->
3
27:1
11-50
40
4
->
4
27:1
11-50
40
4
->
5
27:1
11-50
55
0
-<
6
27:1
11-50
10
0
->
Condition at the time of delivery: Free application
Step
Instrument
Speed (rpm) Torque
(Ncm)
Pump
Direction of
motor rota‐
tion
27:1
11-1500
55
4
->
5.4.7
Surgical motor in the program
Requirement
The mounted INTRA CL light, hand and contra-angle handpiece must be selected
on the display.
If the mounted handpiece or contra-angle handpiece is not selected, the device asks
for the corresponding input.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.4 Program
38/66