Rated power of the trans‐
mitter in W
150 kHz to 80 MHz
d=1.17
P
80 MHz to 800 MHz
d=1.17
P
800 MHz to 2.5 GHz
d=2.33
P
0,01
0,12
0,12
0,23
0,1
0,37
0,37
0,74
1
1,17
1,17
2,33
10
3,70
3,70
7,37
100
11,70
11,70
23,30
For transmitters whose maximum rated power is not in the above table, the recom‐
mended safe distance d in meters (m) can be calculated using the equation for the
respective gap, where P is the maximum rated power of the transmitter in Watts (W)
according to the manufacturer's information.
NOTE 2: These guidelines may not be applicable in every case. The spread of
electromagnetic waves is absorbed and reflected by buildings, objects and people.
Comment 1: To calculate the recommended safe distance from transmitters with a
frequency range of 80 MHz to 2.5 GHz, an additional factor of 10/3 was used to
reduce the probability that a mobile unintentionally brought into the patient area
would cause malfunction.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
9 Details on electromagnetic compatibility | 9.4 Recommended safe distance between portable and mobile
HF telecommunications equipment and the INTRAsurg 300
66/66