Outside Germany

Note
KaVo is not liable for damage arising from transportation.

Immediately inspect the delivery after receipt!

If the packaging is visibly damaged on delivery, please proceed as follows:

1. The recipient of the package must record the loss or damage on the delivery

receipt. The recipient and the representative of the shipping company must sign

this delivery receipt.

Without this evidence, the recipient will not be able to assert a claim for damages

against the shipping company.

2. Leave the product and packaging in the condition in which you received it.

3. Do not use the product.

If the product is damaged but there was no discernable damage to the packaging

upon delivery, proceed as follows:

1. Report the damage immediately or at least 7 days after the delivery to the de‐

livery company.

2. Leave the product and packaging in the condition in which you received it.

3. Do not use a damaged product.

Note
Failure on the part of the recipient to comply with any of the above obligations will

mean that the damage will be considered to have arisen following delivery (in

accordance with CMR law, Chapter 5, Art. 30).

1.5.3

   

Information on the packaging: Storage and transportation

Note
Please keep the packaging in case you need to return the product for servicing or

repair.

The symbols printed on the outside are for transportation and storage, and have the

following meaning:

Transport upright with the arrows pointing upwards

Fragile - protect against impact.

Protect from moisture.

Permissible stacking load

Temperature range

 

 

 

 

 

Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
1 User instructions | 1.5 Transportation and storage

5/66

Содержание INTRAsurg 300

Страница 1: ...Instructions for use INTRAsurg 300 INTRAsurg 300 plus Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...0 plus 13 3 2 Controls 14 3 3 Multifunctional foot control 15 3 4 Rating plate for 300 300 plus 16 3 5 Technical data for the 300 300 plus 16 4 First use 18 4 1 Electrical connection 18 4 2 Unpacking...

Страница 4: ...49 6 2 1 Automated disinfection 49 6 2 2 Manual disinfection 49 6 2 3 Drying 50 6 2 4 Service inspection and testing after preparation 50 6 3 Sterilisation in a steriliser in compliance with DIN EN 13...

Страница 5: ...ly and guarantees zero defects in the material or processing for a period of 12 months from data of purchase subject to the following conditions Upon justified complaints of flaws or a short delivery...

Страница 6: ...packaging is visibly damaged on delivery please proceed as follows 1 The recipient of the package must record the loss or damage on the delivery receipt The recipient and the representative of the sh...

Страница 7: ...e damage immediately or at least 7 days after the delivery to the de livery company 2 Leave the product and packaging in the condition in which you received it 3 Do not use a damaged product Note Fail...

Страница 8: ...Humidity Air pressure Instructions for use INTRAsurg 300 INTRAsurg 300 plus 1 User instructions 1 5 Transportation and storage 6 66...

Страница 9: ...ON CAUTION indicates a hazardous situation that can cause damage to property or mild or moderate physical harm WARNING WARNING indicates a hazardous situation that can cause death or serious injury DA...

Страница 10: ...mmended annual servicing must be done Yearly safety inspections are required KaVo recommends an annual service check In this service check the safety checks are performed according to IEC 62353 2007 a...

Страница 11: ...in ternal components can increase transmission or reduce the electromagnetic im munity of the product Only use accessories recommended by KaVo CAUTION Interactions between devices Effect on proper us...

Страница 12: ...ted sterilized according to the section Preparation methods Additional information can be obtained from KaVo www kavo com or your dental supplier For final disposal contact Germany To return an electr...

Страница 13: ...enings from penetration of liquids Remove liquids from the inside of the device CAUTION Unintended penetration of liquids Electrical shock Do not immerse the product in a tub like container Check and...

Страница 14: ...truction of electronic components Before contacting the hose coupling touch the bottle holder with your hand to release static electricity KaVo recommends that only original KaVo parts be used for ope...

Страница 15: ...functional foot control The KaVo INTRAsurg 300 plus has an additional lighting function and recognition function for the INTRA CL light hand and contra angle handpiece Note Additional information on s...

Страница 16: ...Para meter Open setting mode And Use the parameter button to select changeable parame ters Press the parameter button until the asterisk flashes at the desired position Changeable parameters max torqu...

Страница 17: ...cing the individual steps Press briefly proceed forward press for a long time proceed backwards Double click open selection An acoustic signal confirms proceeding forward backward when the motor start...

Страница 18: ...ctured XXXXXXX Serial number REF Material number Type Device type For disposal information see use in accordance with intended purpose GOST R certification 3 5 Technical data for the 300 300 plus Widt...

Страница 19: ...length hose for the multifunc tional foot control Only available with the INTRAsurg 300plus 3 5 m Standard motor hose length Mat no 1 001 2651 2 00 m Special length motor hose Mat no 1 004 6825 3 0 m...

Страница 20: ...into the device s socket to connect to the mains Both lines L and N are individually fused with T 3 15 H The network frequency can be 50 or 60 Hz Requirement Make sure that the supply voltage is corre...

Страница 21: ...connected to the device Pull out the INTRAsurg 300 300 plus vertically upward and place on a flat sur face 4 3 Adjust the mains input voltage CAUTION Voltage conducting parts Electrical shock Unplug t...

Страница 22: ...ack in the drawer Insert the drawer in the device A code protects the drawer from being incorrectly inserted 1 2 3 4 Instructions for use INTRAsurg 300 INTRAsurg 300 plus 4 First use 4 3 Adjust the ma...

Страница 23: ...g danger from excessively heavy coolant reservoir Malfunctions Do not use coolant reservoirs that can hold more than 1 litre Ensure that the device is standing securely In the case of glass bottles in...

Страница 24: ...uids must be sterilised Plug in the surgery motor at these surgery hose coupling and tighten with a union nut Follow the separate instructions for use for the motor Place the surgery motor on the moto...

Страница 25: ...echanics by lifting the lock Note Insert the pump hose into the pump so that it is not clamped or pinched by the lock Run all hoses relaxed without tension Shove the hoses firmly on the plug in nipple...

Страница 26: ...lf adhesive cable clip Mat no 1 001 3348 to the INTRAsurg 300 300 plus for the location see the illustration Insert the coolant hose into the cable clip and close Note Always monitor the level of the...

Страница 27: ...e device is not monitored we recommend turning it off for safety reasons and to save energy Note The last saved values are displayed 5 2 Surgical motors CAUTION Incorrectly set parameters Damage from...

Страница 28: ...ransferred Check all input values before use Mount the handpiece or angle piece 1 CAUTION Heat transferred from high pressure Burns from touching the high pressure lamp Only operate the motor with the...

Страница 29: ...emove the coolant hose from the handpiece or angle piece Twist the handpiece or contra angle handpiece slightly while pulling it off See also 5 1 1 Start up Page 25 5 2 4 Start the motor Start the mot...

Страница 30: ...r ratio of the handpiece and contra angle handpiece max speed Direction of motor rotation Coolant pump setting max torque Para meter Press the parameter button A star flashes next to the alterable par...

Страница 31: ...Drilling cutting 1 1 3 610 N 1 1 300 40 000 5 5 Grinding Drilling Thread Implant 12 1 3624 N 4 1 C 3 3 1 25 3 300 40 Grinding Drilling Thread Implant 16 1 3624 N 4 1 67 RIC 4 1 18 2 500 45 Grinding Dr...

Страница 32: ...again to start the motor Checking the handpieces and contra angle handpieces Check the set transfer ratio and let the motor and handpiece or contra angle hand piece run for 20 seconds at maximum speed...

Страница 33: ...e application 5 3 4 Set the coolant flow Para meter Press the parameter button until the asterisk flashes at the coolant flow The flow can be changed in the parameter menu by using the Plus and Minus...

Страница 34: ...functional foot control and save it with the enter button 5 3 5 Selecting the handpiece and contra angle handpiece Press the parameter button until the asterisk flashes at the transfer ratio Use the P...

Страница 35: ...i clock wise signal W Y Z Move the shift pedal in the direction of the arrow With counterclockwise rotation the rotational direction display flashes and two acoustic signals sound When the motor is st...

Страница 36: ...5 3 8 Surgical motor in free use Note This function is only supported by the INTRAsurg 300 plus in conjunction with the INTRA CL light hand and angle pieces The recognition function of the INTRA CL li...

Страница 37: ...am The product has a basic program with six programmable steps The current step is shown in the display Press the right button briefly or PRG Press the program button once An acoustic signal confirms...

Страница 38: ...entry 5 4 4 Set parameters Select the step to be changed Para meter Switch to the parameter menu by pressing the parameter button See also 5 3 Free use Page 28 The values shown for a step are default...

Страница 39: ...use 5 4 5 Limit the number of steps Six steps are saved upon first start up and after setting up the delivery status Example Limiting the steps The number of steps should be limited to four Select ste...

Страница 40: ...y Condition at the time of delivery Programme Step Instrument Speed rpm Torque Ncm Pump Direction of motor rota tion 1 27 1 11 1500 55 4 2 27 1 11 50 40 4 3 27 1 11 50 40 4 4 27 1 11 50 40 4 5 27 1 11...

Страница 41: ...gle handpieces without the moun ted and peace and contra angle handpiece Para meter Press the parameter button until the asterisk flashes at the transfer ratio Set the correct parameters with the Plus...

Страница 42: ...piece The transfer ratio and associated values are shown on the display The user can perform additional calibration with the INTRAsurg Calibration 1 002 3569 27 1 angle piece is required Display of th...

Страница 43: ...g input Three acoustic warning signals sound and the read in instrument flashes on the device display with a ques tion mark for example 27 1 Note For safety reasons the motor does not start safety cir...

Страница 44: ...ight only shines when actuated spotlight function Default light when the motor starts Move the shift pedal to the left The light is turned on or off When the light is turned on the following appears i...

Страница 45: ...or restarts under a minimum load of 5 Ncm 5 5 4 Request maximum torque program See also 5 4 8 Request maximum torque Page 39 5 5 5 Recommended programming when setting a number of im plants one after...

Страница 46: ...23 24 25 26 27 28 29 30 correc tion 0 8 0 7 0 6 0 5 0 4 0 3 0 2 0 1 0 0 1 0 2 0 3 0 4 0 5 0 6 0 7 A sterile instrument can also be used for calibration To prevent contamination use a sterile catch pi...

Страница 47: ...n the following appears on the display If calibration is unsuccessful the following appears on the display Note The output values are retained in case of interruption or error The calibration automati...

Страница 48: ...ch as saline coolant residue blood and saliva with a disposable cloth immediately after use Only use fully functional products for treating patients Observe all current national hygiene regulations Se...

Страница 49: ...loth on edges and press so that liquid will not penetrate Remove all residue and dry the device Note After uses that run the risk of infection rinse the coolant lines of the handpieces and contra angl...

Страница 50: ...d Observe the necessary measures for infection protection W Y Z Press left button 3 sec The rinsing function is selected See also 5 3 7 Rinsing function Page 33 The rinsing function is shown in the di...

Страница 51: ...pplicable for INTRAsurg device parts 6 2 2 Manual disinfection Pull the insertion needle out of the coolant reservoir Open the lock and remove the hose Remove all hoses marked gray including tips and...

Страница 52: ...chment of hose connections Check control commands on foot control Check recognition function of INTRAsurg 300 plus Comply with hygiene requirements sterility during the check If there are any sites of...

Страница 53: ...icrobiologically appropriate coolant liquid must be used Only use prescribed isotonic saline solution E NaCl 0 9 identified as an infusion solution follow the instructions on the package insert for co...

Страница 54: ...eam at 135 1 gravity for at least 4 minutes Note Let the sterilised materials cool and try to the ambient temperature before using them again 6 3 3 Storage Observe all necessary measures for hygiene w...

Страница 55: ...not plugged in Plug in the mains power line The wrong handpiece or angle piece is mounted or set on the de vice Adjust the right handpiece or an gle piece on the device Unknown Turn the machine off an...

Страница 56: ...ed or set on the de vice Adjust the right handpiece or an gle piece on the device The motor makes a grinding noise The motor is not correctly plugged on or screwed on Firmly insert the motor hose into...

Страница 57: ...connec ted to the motor hose Properly plug in the motor Display remark 67 motor not plugged When the motor starts the discon nection of the motor from the hose is detected Plug the hose onto the moto...

Страница 58: ...drain the lines remove the insertion needle from the saline solution res ervoir and activate the pump by using the multifunctional foot control until the hoses are free of saline solution Remove the...

Страница 59: ...Pump hose short less accumulated volume in the coolant hose less potential dripping Pump hose long large accumulated volume in the coolant hose increased drip ping possible Observe hygiene to prevent...

Страница 60: ...5 5188 cut section off the cou pling Mat no 1 001 6462 and the handpiece or angle piece and exchange 7 2 2 Exchange the coolant hose in the motor hose Note Run the new hose straight in the direction o...

Страница 61: ...can be ordered under Mat no 1 001 9902 pack age contents 10 units This should be used for applications with an increases infection risk or when the motor hose cannot be prepared according to the exis...

Страница 62: ...completely open the inserted hose clip The coolant flow is also affected by the outlet in the handpiece or angle piece A focused water jet should arise Set the pump flow on the device Note To avoid dr...

Страница 63: ...ame __________________ Step Instrument Speed rpm Torque Ncm Pump Motor r l Free Instrument Speed rpm Torque Ncm Pump Motor r l Date _______________________ Name __________________ Step Instrument Spee...

Страница 64: ...o 0 726 2922 IPX 8 foot control element with 3 5 m connection cable please in quire Mat no 1 004 8276 Steri Set INTRAsurg 300 please inquire Mat no 1 001 4953 includes the fol lowing parts 1x motor ho...

Страница 65: ...esidential ones and fa cilities that are directly connected to a public power supply that also sup plies residential buildings Transmissions of voltage fluctua tions or flicker according to IEC 61000...

Страница 66: ...EC 61000 4 11 5 UT for 1 2 period 95 interruption 40 UT for 5 periods 60 interruption 70 UT for 25 periods 30 interruption 5 UT for 5 s 95 interruption 5 UT for 1 2 period 95 interruption 40 UT for 5...

Страница 67: ...ers such as base stations of mobile tele phones and land radio devices amateur radio stations AM and FM radio and tel evision broadcasters cannot be theoretically predetermined To determine the elec t...

Страница 68: ...o the manufacturer s information NOTE 2 These guidelines may not be applicable in every case The spread of electromagnetic waves is absorbed and reflected by buildings objects and people Comment 1 To...

Страница 69: ......

Страница 70: ...1 001 3529 Fk 2010506_09 en...

Отзывы: