4 Installation steps
30
FOCUS
4.8.3 Using terminal strip for connecting the
control panel
The unit includes remote control panel cable with
RJ-45 connector at both ends. In case it is not
possible to pull the RJ-45 connector through
conduits at the place of installation the following
instructions may be used.
In this instruction the terminal strip (X61) at the
unit connection box board is used instead of the RJ-
45 terminal connector (X62 or X63).
NOTICE!
Use only the original cable that has been
delivered by the manufacturer for installing the
remote control panel
There are two different types of cables delivered,
with different color coding. Check the RJ-45 pin
colors and determine cable type used. (see tables
below).