4 Installation steps
40
FOCUS
See section 3 Pre-installation for the mains voltage
and fuse requirements.
Use only approved cables and plugs.
Mains connector type:
NEMA 5-15P (115 V), NEMA 6-15P (230 V) or
similar Hospital Grade
Power supply cord type:
EU 230V:
H05VV-F / 1.5 mm
2
(UL 2587), max.
3 m / 9.8 ft.
115V:
115 V / 15A, Type: SJT, min. 14AWG,
max. 3 m / 9.8 ft. Complies with UL817 and
CSA C22.2 NO 21.
WARNING!
Only a qualified technician is permitted
to carry out these connections.
WARNING!
Before connecting to the mains, make
sure that the cable connections to the fuses are
wired as instructed.
4.10.2 Converting a non-permanent (plug)
wiring to permanent
NOTICE!
For permanent installation, a separate
lockable mains switch (not supplied) is required to
be installed.
1.
Remove the cable clamp and the power supply
cable.
Содержание FOCUS 215720
Страница 1: ...FOCUS Intraoral X ray Installation Manual 215720 rev 11 0 805 4950 ENGLISH Dental Excellence FOCUS...
Страница 2: ......
Страница 4: ......
Страница 6: ...ii...
Страница 12: ...2 Unit description 6 FOCUS Exposure time control...
Страница 44: ...4 Installation steps 38 FOCUS 4 9 4 Example 4 Remote control and external exposure button in series...
Страница 55: ...6 Dimensions FOCUS 49 6 Dimensions Wide wall mount plate measures in mm...
Страница 56: ...6 Dimensions 50 FOCUS Narrow wall mount plate measures in mm...
Страница 57: ...6 Dimensions FOCUS 51 Remote control frame measures in mm...
Страница 58: ...6 Dimensions 52 FOCUS Wall mount installation dimensions...
Страница 59: ...6 Dimensions FOCUS 53...
Страница 60: ...6 Dimensions 54 FOCUS...
Страница 61: ...6 Dimensions FOCUS 55...
Страница 62: ...6 Dimensions 56 FOCUS OpenMount installation dimensions...