3 Pre-installation
FOCUS
7
3 Pre-installation
3.1 Pre-installation checklist
Radiation protection:
Follow all local regulations for radiation shielding of
the X-ray room.
During the exposure the operator must follow
applicable radiation shielding requirements and
stay at least 7 feet (2 meters) from the radiation
source.
Mounting requirements:
The wall and mounting hardware must sustain a
pull of 350 lbs./1500N, from each wall bracket bolt.
When replacing an old X-ray unit using a wide wall
mount plate, observe the space that is covered by
the wall mount plate to minimize painting where
applicable.
Mounting on a treatment unit is allowed only when
the manufacturer of the treatment unit has tested
the mechanical stability and allow the integration.
For KaVo treatment unit integration refer the
compatibility table of the accessories available.
For OpenMount installation, the mounting shall
contain a bearing designed for a shaft of diameter
of 32 mm (h7) and a length of 105 mm when
mounting on a 3rd party adapter, to ensure that
the maximum load ratings of the adapter are not
exceeded by the FOCUS OpenMount unit. The
installation must provide tilt adjustment to level
and balance the arm. The installation must not
allow rotating the X-ray unit more than 360
degrees to prevent twisting of the cables.
Electrical requirements:
Mains voltage:
Verify the mains voltage (3 wire: single-phase, 115
V or 230 V with protective ground).
Fuses:
Содержание FOCUS 215720
Страница 1: ...FOCUS Intraoral X ray Installation Manual 215720 rev 11 0 805 4950 ENGLISH Dental Excellence FOCUS...
Страница 2: ......
Страница 4: ......
Страница 6: ...ii...
Страница 12: ...2 Unit description 6 FOCUS Exposure time control...
Страница 44: ...4 Installation steps 38 FOCUS 4 9 4 Example 4 Remote control and external exposure button in series...
Страница 55: ...6 Dimensions FOCUS 49 6 Dimensions Wide wall mount plate measures in mm...
Страница 56: ...6 Dimensions 50 FOCUS Narrow wall mount plate measures in mm...
Страница 57: ...6 Dimensions FOCUS 51 Remote control frame measures in mm...
Страница 58: ...6 Dimensions 52 FOCUS Wall mount installation dimensions...
Страница 59: ...6 Dimensions FOCUS 53...
Страница 60: ...6 Dimensions 54 FOCUS...
Страница 61: ...6 Dimensions FOCUS 55...
Страница 62: ...6 Dimensions 56 FOCUS OpenMount installation dimensions...