3 Pre-installation
FOCUS
9
NOTICE!
The reaches listed are with the 9-inch
(229 mm) cone. Using the optional 12-inch (305
mm) cone will reduce the arm reach by 3 inches
(76 mm).
3.2 Tools and hardware
NOTICE!
All screws, bolts, nuts and washers used
in the unit are in metric sizes.
The following tools are required during the
installation:
Metric Allen wrench set
Electric drill, drill bits (8 mm & 10 mm or equal)
Mounting hardware for wall bracket
Pencil or drawing ink
Spirit Level
Hammer
Anchor driver (concrete wall)
Pozidrive or phillips screwdrivers (PZD-1 and
PZD-2)
Pliers
Knife or scissors
Wrenches, 13 mm and 19 mm
Tape measure
Make sure that you have all tools and items needed
prior to installation.
NOTICE!
This unit is intended for fixed installation
only, therefore mobile installation is prohibited.
Содержание FOCUS 215720
Страница 1: ...FOCUS Intraoral X ray Installation Manual 215720 rev 11 0 805 4950 ENGLISH Dental Excellence FOCUS...
Страница 2: ......
Страница 4: ......
Страница 6: ...ii...
Страница 12: ...2 Unit description 6 FOCUS Exposure time control...
Страница 44: ...4 Installation steps 38 FOCUS 4 9 4 Example 4 Remote control and external exposure button in series...
Страница 55: ...6 Dimensions FOCUS 49 6 Dimensions Wide wall mount plate measures in mm...
Страница 56: ...6 Dimensions 50 FOCUS Narrow wall mount plate measures in mm...
Страница 57: ...6 Dimensions FOCUS 51 Remote control frame measures in mm...
Страница 58: ...6 Dimensions 52 FOCUS Wall mount installation dimensions...
Страница 59: ...6 Dimensions FOCUS 53...
Страница 60: ...6 Dimensions 54 FOCUS...
Страница 61: ...6 Dimensions FOCUS 55...
Страница 62: ...6 Dimensions 56 FOCUS OpenMount installation dimensions...