5 Finalizing the installation
48
FOCUS
5.3 System checks before use
1.
Check that the device does not move and that
the arm is not drifting.
2.
Check the local regulations on radiation safety.
Carry out the required radiation measurements
and inform the local authorities.
3.
Move the tube head to up and down to check
that the movements are smooth to both
directions and that the tube head stays in
position.
4.
Check Protective Earth (PE) grounding
resistance according to the IEC 62353 (Medical
electrical equipment - Recurrent test and test
after repair of medical electrical equipment).
5.
Check visually that the exposure indicator light
illuminates for the duration of the exposure.
6.
Confirm that the audible indicator beeps for the
duration of the exposure.
7.
Check all functions of the remote control.
Содержание FOCUS 215720
Страница 1: ...FOCUS Intraoral X ray Installation Manual 215720 rev 11 0 805 4950 ENGLISH Dental Excellence FOCUS...
Страница 2: ......
Страница 4: ......
Страница 6: ...ii...
Страница 12: ...2 Unit description 6 FOCUS Exposure time control...
Страница 44: ...4 Installation steps 38 FOCUS 4 9 4 Example 4 Remote control and external exposure button in series...
Страница 55: ...6 Dimensions FOCUS 49 6 Dimensions Wide wall mount plate measures in mm...
Страница 56: ...6 Dimensions 50 FOCUS Narrow wall mount plate measures in mm...
Страница 57: ...6 Dimensions FOCUS 51 Remote control frame measures in mm...
Страница 58: ...6 Dimensions 52 FOCUS Wall mount installation dimensions...
Страница 59: ...6 Dimensions FOCUS 53...
Страница 60: ...6 Dimensions 54 FOCUS...
Страница 61: ...6 Dimensions FOCUS 55...
Страница 62: ...6 Dimensions 56 FOCUS OpenMount installation dimensions...