5 Finalizing the installation
FOCUS
47
5 Finalizing the installation
5.1 Fastening the covers
Place the small U-shaped cover on top of the
adjustment part.
If the wire from the remote control is surface
mounted, remove a piece from the mounting
bracket cover in order to allow the remote control
cable to fit in. If the unit is supplied with two
remote controls, do the same on the opposite side.
Fasten the cover with two screws at the bottom.
5.2 Calibration to be done before use
Calibrate the unit by exposing manually 60 kV,
1.00 s exposures 8 times. This calibrates filament
preheat and tube current.