![KaVo ELECTROmatic TM Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/kavo/electromatic-tm/electromatic-tm_instructions-for-use-manual_3703148048.webp)
Instructions for use ELECTROmatic TM and TMM/TMC
4 Installation | 4.8 Connecting the ELECTROmatic
48 / 90
CAUTION
Damage to the dentist element.
Installations involving an intervention on the treatment unit might damage
components, which can interfere with the safe function and cause injury.
▶
Have installations involving an intervention on the treatment unit per-
formed by trained expert personnel only.
▶
Have the treatment unit subjected to a safety check after installation.
The following parts from the scope of delivery and the installations set are
needed:
▪ 1x Insert holder
▪ 1x Mounting bracket
▪ 1x Cover panel
▪ 4x + 2x screws M4x12
④
with self-locking nuts
①
▪ 4x + 2x washers
②
▪ Optional: 4x spacer bolts, 35 mm
⑦
▪ Optional: 1x mounting plate
▪ 2x plastic screws
⑥
▶
Open the cabinet.
▶
Define the position of the control panel in the cabinet or on the wall and
draw a hole pattern for attachment. If possible, use existing screws or per-
forations as screw positions.
▶
Installation on a cabinet: Use 2 plastic screws to screw the control panel to
the cabinet.
▶
Installation on the wall: Use 2 screws
④
and washers
②
to screw the angle
bracket with control panel to the wall and secure it with the nuts
①
.
▶
Mount the control unit.
See also:
2
4.4 Installation position 1: Mount below a holder, Page 36
4.8 Connecting the ELECTROmatic
See also:
2
5.1.2 Connecting the ELECTROmatic, Page 49
2
5.1.5 Connect the power supply, Page 51
4.9 Check the installation
▶
Check if the fastening is secure.
▶
Perform the startup.
See also:
2
Содержание ELECTROmatic TM