![KaVo ELECTROmatic TM Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/kavo/electromatic-tm/electromatic-tm_instructions-for-use-manual_3703148051.webp)
Instructions for use ELECTROmatic TM and TMM/TMC
5 Startup | 5.1 Connector
ELECTROmatic TMM/TMC
ELECTROmatic TM
▶
Set the spray regulation on the hose to maximum amount of water.
See also:
2
6.3 Regulating the spray water, Page 56
5.1.5 Connect the power supply
CAUTION
Power supply and cords/hoses are on the floor.
Slipping and tripping.
▶
Power supply and cord/hoses need to be set-up/routed appropriately such
that they are not on the floor.
NOTICE
Property damage from non-approved power supply.
Property damage to the product.
▶
Operate the product with the type 4882 power supply (Mat. no.
1.005.0120) exclusively.
Note
The connection of the power supply must comply with the country-specific
regulations and requirements for medical devices.
Note
The power supply automatically adjusts to the available mains voltage.
Note
The protective earth conductor is used as functional earthing (FE) rather than
as protective earthing (PE).
Note
Use the power supply in dry environments exclusively. Make sure that the
power supply is protected from the ingress of liquids.
▶
Connect the type 4882 power supply to the socket of the ELECTROmatic.
▶
Connect the power cable first to the power supply and then to the supply
mains socket.
▶
Route the cables appropriately such that there is no kinking and, if possible
and permitted by the dental unit, in the support arm.