Instructions for use ELECTROmatic TM and TMM/TMC
8 Reprocessing steps in accordance with DIN EN ISO 17664 | 8.3 Packaging
76 / 90
▶
Connect the ELECTROmatic to the dental unit.
▶
Follow the instructions for germ reduction of the dental unit.
The product should be used according to the instructions of the manufacturer
and the Instructions for Use of the dental unit.
8.2.3 Automated external and internal disinfection
The exterior and interior of this product are not designed for automated disin-
fection.
8.3 Packaging
Not applicable.
8.4 Sterilisation
Not applicable.
8.5 Storage
Prepared products must be stored appropriately in a dry, dark, cool room such
that they are protected from germs (as far as possible) and dust.
Note
Comply with the expiry date of the sterilised items.
8.6 Service, inspection and testing after preparation
Note
It is essential to comply with the hygiene requirements (sterility) during the
test after reprocessing. If sites of fracture and clear changes of the surface
are visible, the parts need to be checked by the Service.
Check for cleanliness, intactness, servicing, and repair as described in the fol-
lowing:
▶
Check the adjustable functions of the unit and the motor function.
▶
Check the control commands on the foot control.