![KaVo ELECTROmatic TM Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/kavo/electromatic-tm/electromatic-tm_instructions-for-use-manual_3703148044.webp)
Instructions for use ELECTROmatic TM and TMM/TMC
4 Installation | 4.7 Installation position 4: Mount control panel as remote control
44 / 90
4.7.2 Mount control panel on a holder / on the backside
of a holder
CAUTION
Damage to the dentist element.
Installations involving an intervention on the treatment unit might damage
components, which can interfere with the safe function and cause injury.
▶
Have installations involving an intervention on the treatment unit per-
formed by trained expert personnel only.
▶
Have the treatment unit subjected to a safety check after installation.
The following parts from the scope of delivery and the installations set are
needed:
▪ 1x Insert holder
▪ 1x Mounting bracket
▪ 1x Cover panel
▪ 4x + 2x screws M4x12
④
with self-locking nuts
①
▪ 4x + 2x washers
②
▪ Optional: 4x spacer bolts, 35 mm
⑦
▪ Optional: 1x mounting plate
▪ 2x plastic screws
⑥
▶
Define the position of the control panel on the holder or on the rear of the
holder and draw a hole pattern for attachment. If possible, use existing
screws or perforations as screw positions.
▶
Drill the holes.
▶
Screw the mounting bracket with control panel to the holder or to the rear
of the holder using the 2 screws
④
, washers
②
and nuts
①
.