5 Commissioning
Units
Select from list ↑↓
Min Value
Min. process variable (PV) value that corresponds
to minimum frequency
Max Value
Max. process variable (PV) value that
corresponds to maximum frequency
Damping
Additional smoothing of the current output, the
higher the damping factor, 1 ... 255 s
Error
Defines output behaviour in the event of error
Select from list ↑↓
Relay /
Optical relay
Mode
Off –
Permanently off
On –
Permanently energised
Alarm –
PV alarm switch
Math
– Calculated value alarm switch
Fault –
Allocated to system failures, see error
report list
On Point
Value of PV at which the relay energises when in
alarm mode
Off Point
Value of PV at which the relay de-energises when
in alarm m ode
Current In
Source
(channel)
Select from list ↑↓
Off, Channel 1, Channel 2, Math 1, Math 2
System, Test
Source
(value)
Select from list ↑↓
Density, Viscosity, Temperature, Other
Minimum, Maximum, Span settings as on outputs
PT100
Temperature inputs
Source
Fixed –
A fixed temperature can be entered under
value
PT100 –
Value read from PT100 temperature
sensor in ºC
Value
Enter fixed user defined value
0 ... 250 ºC
Offset
Enter fixed user defined value
-100 ... 100 ºC
RS 485
[where specified]
Modbus RTU
[where specified]
HART
[where specified]
System
Instrument
info
Model Code
100, 101 or KFPROG for programmer
Serial No.
Example:
10100026
[screen]
HW Revision
Hardware revision. Example:
1.1, 1.2
SW Revision
Software revision. Example:
2.3, 1.4
[programmer]
HHP HW Rev
Programmer hardware eg. 2.0
HHP SW Rev
Programmer software eg. 3.9
XMTR HW Rev Flowmeter hardware eg. 2.0,1.5
XMTR SW Rev Flowmeter software eg. 3.9,2.4
Calculation
Low F Cut
± Low flow velocity cut off
0 ... 0.025 m/s
Max F Cut
± Maximum flow velocity cut off
31