KATflow 100
5 Commissioning
0 ... 30 m/s
Corrected
Apply flow velocity profile correction
Yes, No
PV Offset
Calibration process variable zero offset
-30 ... 30 units
PV Scaling
Calibration process variable gradient scaling
0 ... 10000 units
Zero Cal
Zero calibration settings
Zero
Perform auto zero calibration
Yes, No
Track
Track zero offset
Yes, No
Delta
Zero flow delta time offset in ns, read from sensor
PROM or entered directly for special sensors
Timeup
Upstream transit-time offset in µs, allows for fixed
delays in special sensors, buffer rods and
extension leads
Heat Capacity
Specific heat capacity of medium
User
Identifier
Example:
Pump P3A
9 character string
Tag No.
Example:
1FT-3011
9 character string
Test
Installation
Control system simulation: 60 second increase of
flow velocity in m/s from 0 to programmed Max F
Cut, then 60 second decrease, i.e. the process
variable changes over complete possible range.
All configured outputs exhibit their programmed
behaviour.
Yes / No
Test modes also available for display, keypad,
memory. peripheral and ultrasonic components
Settings
Date, Time,
Date Format
Enter or select from list
Language
Select from list ↑↓(as available)
English, German, French, Spanish, Russian
Keypad
Enable keypad sound
Yes / No
Defaults
Reload factory default settings, except date/time
Yes / No
Diagnostics
[where specified]
Datalogger
[where specified]
Interval
Enter logging interval in seconds ('0' for off,
0..999s)
Selection
Select up to 10 items from list ↑↓
m/s, f/s, in/s,
m3/h, m3/min, m3/s, l/h, l/min, l/s,
USgall/h, USgall/min, USgall/s,
bbl/d, bbl/h, bbl/min,
g/s, t/h, kg/h, kg/min,
m3, l, Usgall, bbl, g, t, kg
W, kW, MW, J, kJ, MJ,
Sig dB (signal), noise dB, SNR,
C m/s (sound speed), CU (housing temperature)
Tin, Tout (inlet and outlet temperature)
32