KATflow 100
4 Operation
(Screen)
Use cursor keys to select
Start Measurement
.
Confirm by pressing <
ENTER
>.
Sensor positioning screen: Mount transducers
with suggested spacing and use middle bar for
fine adjustment of position (central position is
desired). Observe signal-to-noise (upper bar)
and quality (lower bar). These should be of
identical length.
Confirm by pressing <
ENTER
> to obtain
measurements (screen) or to return to menu
(programmer).
Note : Numbers shown are for indication
only.
(Programmer)
Use cursor keys to select
Start/Stop
. Confirm
by pressing <
ENTER
>.
Programmer will send the start command to
the flowmeter. When complete, process
outputs and local display (if fitted) will be
active.
Programmer screen will revert to this display.
Success!
4.4 Measurements
4.4.1 Main process value (PV) – (screen only)
Measurement is started using the Quick Setup Wizard. Once all the parameters are
programmed, any subsequent power-on sequences will immediately give the main
PV as a display (where specified) and/or as an output as appropriate.
Display screen
Operation
The main process value can be changed
in the menu structure.
Press
<ESC>
at any time to access the
main menu.
Change to other measurement and
diagnostic screens by pressing the arrow
keys
(where fitted).
26
Measurement
screens