28
Bipolar Electrodes/
Coagulation Probes
Indication
•
To coagulate tissue via a high frequency
electrical current waveform during
general, endoscopic and laparoscopic
surgical procedures.
Inspection
•
The shaft of the instruments should be
straight.
•
The insulation should be intact. There
should be no dents, scratches, burn
marks, cracking or splitting of the
insulation.
Cleaning
•
Follow the general instructions for cleaning
indicated in the Cleaning Section of this
manual.
Sterilization:
WARNING:
High-level disinfection is NOT
recommended for “critical” instruments to
be used for laparoscopic, hysteroscopic,
gynecology or neuroendoscopy procedures.
•
All electrodes/coagulating probes may
be ethylene oxide, or steam sterilized.
Please consult the Sterilization Section
of this manual for complete instructions
for validated sterilization procedures.
Caution:
The following bipolar electrodes
are sterile, single-use instruments and must
not be reprocessed. Dispose of the electrode
after use.
- 28164FGB-S
- 28164FGL-S
- 28164FGC-S
- 28164FGM-S
- 28164FGD-S
- 28164FDG-S
- 28164FGK-S
- 28164FDN-S
Содержание TAKE-APART