WARNING:
DO NOT use laser fibers when
using electrosurgical instruments.
WARNING:
Avoid long activation of the
electrosurgical generator to reduce the pos-
sibility of RF leakage current.
WARNING:
Use surgical gloves specifically
designated by their manufacturer for
electro
-
surgical procedures. The use of
other gloves may result in electrosurgical
shocks or burns to the surgeon.
WARNING:
If a reduction in the expected
electrosurgical output is noted, check the
connections to ensure proper seating.
Check the active electrode insulation for
possible damage before increasing the
power of the electrosurgical generator.
WARNING:
Keep the distal tip of any
bipolar instrument in their field of view at all
times.
WARNING:
Electrodes and probes of
monitoring, stimulating and imaging devices
can provide paths for high frequency currents
even if they are battery powered, bipolar or
isolated at 60 Hertz (Hz).
CAUTION:
When removing the cord from
the electrosurgical unit, firmly grasp and
pull the connector. Do not pull the cord, the
cord may be damaged or broken.