Light source – troubleshooting
Symptom
Possible causes
Remedy
Light source cannot be connected to the MoVeo
videoscope
An adaptor sleeve is screwed
onto the videoscope light port
pin
Remove the screw-on adaptor
sleeve or, depending on the version,
use a suitable sleeve.
6JGNKIJVUQWTEGECPPQVDGEQPPGEVGFVQCP
endoscope or flexoscope
An incorrect adaptor sleeve is
screwed onto the light port pin
or there is no adaptor sleeve
.'&NKIJVUQWTEGDGEQOGUXGT[YCTOCHVGTGZVGPFGF
use.
6JG.'&NKIJVUQWTEGKUQRGTCVGF
for more than 30 minutes at
maximum light intensity
Reduce the light intensity if light
KUUVKNNTGSWKTGF1TUYKVEJVJG.'&
light source off for a few minutes to
allow the housing to cool down.
6JGNKIJVKPVGPUKV[QHVJG.'&NKIJVUQWTEGKUVQQNQYQT
VJG.'&NKIJVUQWTEGFQGUPQVGOKVCP[NKIJV
6JGDCVVGT[KUPQVEJCTIG
enough
Replace the batteries with the
designated fully charged batteries.
Batteries either do not work or
not to a sufficient degree
Check that the batteries are
inserted correctly.
Check that the specified batteries
are used.
Check the battery charge state and
use fully charged batteries.
#FLWUVVJGNKIJVKPVGPUKV[VQCNQYGT
level.
.KIJVEQPGQH.'&NKIJVUQWTEGHNCUJGU
Battery charge state is low
Replace the batteries with the
designated fully charged batteries. If
PGEGUUCT[CFLWUVVJGNKIJVKPVGPUKV[
to a lower level in order to continue
using the light source for a few
minutes and replace the batteries
with fully charged ones as soon as
possible.