Easidew PRO XP User’s Manual
30
97442 Issue 2.5, January 2016
APPENDIX D
Displayed Text
Description
Setting sampling rate St
Press
E
button
Press
and
buttons to change the sampling rate from 1 to 10
seconds
1 = Factory default
Press
E
button. Press
to go to Engineering Unit
Displayed Text
Description
Set-up Engineering Unit
Press
E
button
Press
and
buttons to select the unit
- nonE -
for no unit on the display (factory default setting) see D.5
- °C, °F, K, %
(There is a 6 second cycle
- measured value is displayed for 4 seconds
- the unit is displayed for 2 seconds)
Press
E
button
Exit from menu and save settings:
Press
and
buttons (possible from each page)
D.5
Moisture Scaling Label
The Easidew PRO XP has 3 standard ranges and scales which are as follows:
EA-XP-TX
-110 to +20°Cdp output range
-100 to +20°Cdp output range
EA-XP-LQ-TX
0-3000ppm
W
The EA-XP-TX units will be configured for the range/scale ordered (as above) and will constantly
display the measured value. The display meter will have a stick-on label defining the scale in °C DP.
E
Easidew PRO XP
Moisture Transmitter
°C DP
Label
If ordered with a non-standard range/scale the unit could be set up with the following:
°F DP
lbs/MMSCF
ppm
V
ppm
W
mg/m
3
Alternatively, an Easidew PRO XP Communication Kit (XP-CK) can be used to change the range/scale
and this Kit will be shipped with its own manual and 2 additional sets of scaling labels so that the
label can be changed as required.
Содержание Easidew PRO XP
Страница 28: ...Easidew PRO XP User s Manual Kahn Instruments 19 APPENDIX A Appendix A Technical Specifications...
Страница 31: ...Easidew PRO XP User s Manual 22 97442 Issue 2 5 January 2016 APPENDIX B Appendix B Hazardous Area Certification...
Страница 34: ...Easidew PRO XP User s Manual Kahn Instruments 25 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 35: ...Easidew PRO XP User s Manual 26 97442 Issue 2 5 January 2016 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 41: ...http www kahn com...