Easidew PRO XP User’s Manual
18
97442 Issue 2.5, January 2016
MAINTENANCE
4.4
O-Ring Replacement
Do not touch the filter with bare hands
1.
Identify the O-ring to be removed, as shown below.
BS116 (3/4” x 3/32”) viton,
75 shore
2.
Carefully slide tweezers, thin bladed screwdriver or a blunt needle under
the outer edge of the O-ring.
NOTE: Take care not to scratch any of
the surfaces of the surrounding metal component.
3.
Move the tool around the circumference to assist the extraction process.
Slide the O-ring clear of the thread and filter.
4.
Make sure the groove has no scratches and is free from grease, dirt
or debris. Slide the new O-ring over the filter and thread and into the
groove.
NOTE: Do not touch the filter with bare hands.